Remote telemetry may facilitate the management of implantable devices. We tested the reliability of a new automatic, wireless home monitoring (HM) system that archives data every 24 hours. We retrospectively analyzed archival data from 276 consecutive pacing system implants to define temporal atrial fibrillation (AF) patterns and associated ventricular rate. An "AF day" was defined by a >20%/24 hour mode switch (MS) duration, irrespective of the MS number. Management decisions resulting from transmissions were noted. A pilot study confirmed that 89% of 22,356 transmissions were successful, of which >90% were received in <5 minutes. Data integrity was 100% preserved. Overall, AF developed in 29 patients (10.5%), representing a total of 645 AF days (mean = 22.2 +/- 29.6 AF, median = 9 days), over 12 +/- 2 months of monitoring. AF was infrequent (/=1 day during which the AF burden was >50% of 24 hours. Ventricular rates during 645 AF days in 29 patients averaged 95.1 +/- 9.9 beats/min (median = 94 beats/min). Ventricular rates were >80 beats/min in 25 +/- 30 AF days (median = 11 days). HM enabled rapid anticoagulation decisions. In recipients of implantable devices, automatic wireless telemetry with HM was efficient and reliable. Its application may overcome some current challenges in AF management by early notification and precise measurement of both AF burden and ventricular rate during AF.
机译:远程遥测可以促进可植入设备的管理。我们测试了新的自动无线家庭监控(HM)系统的可靠性,该系统每24小时存档一次数据。我们回顾性分析了来自276个连续起搏系统植入物的档案数据,以定义时间性房颤(AF)模式和相关的心室率。 “ AF天”是由> 20%/ 24小时模式切换(MS)持续时间定义的,而与MS编号无关。记录了由传播产生的管理决策。一项初步研究证实,在22,356次传输中,有89%成功,其中在不到5分钟的时间内收到了90%以上的传输。数据完整性100%得到保留。总体而言,在监测的12 +/- 2个月中,有29例患者(10.5%)出现了AF,代表总共645 AF天(平均= 22.2 +/- 29.6 AF,中位数= 9天)。 80%的患者很少发生房颤( / = 1天,在此期间房颤负担超过24小时的50%。 29例患者在645个AF日内的心室率平均为95.1 +/- 9.9次/分(中位数= 94次/分)。在25 +/- 30 AF天(中位数= 11天)中,心室率> 80次/分钟。 HM使快速抗凝决策成为可能。在可植入设备的接收者中,带有HM的自动无线遥测技术既高效又可靠。它的应用可以通过尽早通知和精确测量房颤期间的房颤负担和心室率来克服房颤管理方面的当前挑战。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号