The AutoCapture (AC) function of new pacemakers (PM) from St Jude Medical (SJM) was originally recommended for use with low polarization (LP) ventricular leads only.However, recent reports have encouraged the use of the AC function with various leads, including those lacking a special LP design. The objective of this study was to analyze the reliability and safety of the AC algorithm application with different types of pacing leads.The study group comprised 30 consecutive patients with AC PMs connected to three different types of non-LP leads. Ten patients with SJM LP leads served as the control group. The study protocol included a complete AC function test using four different pulse widths (PW). The pacing threshold was independently assessed by a manual/semiautomatic check. Erratic behavior of polarization measurements with increasing PWs was demonstrated in 43% (n = 13) of the study group. Invalid polarization measurements resulted in erroneous algorithm recommendation to apply AC function in 17% (n= 5) of the study patients. Subsequent AC function activation lead to incorrect threshold determination due to missed noncapture in three patients. AC function should be applied with caution with non-LP leads. "Off label" use of these leads may cause erroneous polarization signal measurements which, in some cases, may result in incorrect pacing threshold determination, rendering a potential risk to dependent patients.
机译:最初建议将St Jude Medical(SJM)新型起搏器(PM)的AutoCapture(AC)功能仅用于低极化(LP)心室导线。但是,最近的报告鼓励将AC功能用于各种导线,包括那些缺乏特殊LP设计的产品。这项研究的目的是分析使用不同类型的起搏导线的AC算法应用的可靠性和安全性。研究组包括30位连续的AC PMs患者,这些患者与三种不同类型的非LP导线相连。十例SJM LP导联患者作为对照组。研究方案包括使用四个不同脉冲宽度(PW)的完整AC功能测试。起搏阈值通过手动/半自动检查独立评估。在研究组中有43%(n = 13)证明了极化测量随PW增大的不稳定行为。无效的极化测量结果导致错误的算法建议,以在17%(n = 5)的研究患者中应用AC功能。随后的AC功能激活由于三位患者错过未捕获而导致阈值确定错误。对于非LP引线,应谨慎使用交流功能。这些导线的“不合格”使用可能会导致极化信号测量错误,在某些情况下,可能会导致错误的起搏阈值确定,从而给相依患者带来潜在风险。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号