首页> 外文期刊>PACE: Pacing and clinical electrophysiology >Interruption of Pacing Following Nonsustained Ventricular Tachycardia in an AAI Programmed Implantable Cardioverter Defibrillator
【24h】

Interruption of Pacing Following Nonsustained Ventricular Tachycardia in an AAI Programmed Implantable Cardioverter Defibrillator

机译:AAI程序植入式心脏复律除颤器中非持续性室性心动过速后的起搏中断

获取原文
获取原文并翻译 | 示例
           

摘要

BackgroundUnnecessary ventricular pacing from cardiac implantable electronic devices has been associated with long-term risks (heart failure, atrial fibrillation, and possibly stroke). Several device programming strategies are available to minimize ventricular pacing, each potentially associated with unintended consequences. Occasionally, the only effective means is to program to the AAI(R) pacing mode. However, in one manufacturer's implantable cardioverter defibrillators (ICDs), the AAI(R) mode has the potential risk of prolonged pacing cessation following a nonsustained ventricular tachycardia (NSVT).
机译:背景技术心脏植入式电子设备不必要的心室起搏与长期风险(心力衰竭,心房颤动以及可能的中风)有关。有几种设备编程策略可用于最小化心室起搏,每种策略都可能导致意想不到的后果。有时,唯一有效的方法是将程序编程为AAI(R)起搏模式。但是,在一个制造商的植入式心脏复律除颤器(ICD)中,AAI(R)模式具有在非持续性室性心动过速(NSVT)后长时间停止起搏的潜在风险。