Implantable cardioverter defibrillator in a 4-month-old infant with cardiac arrest associated with a vascular heart tumor.

机译:在4个月大的婴儿中植入可植入的心脏复律除颤器,其心脏骤停与血管性心脏肿瘤有关。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A previously healthy male infant was resuscitated after spontaneous ventricular fibrillation at 9 weeks of age. Echocardiography revealed three tumors in the left ventricle not amenable to complete resection. Despite treatment with antiarrhythmic agents the ventricular arrhythmias continued. When the child was 4 months old and weighed 7 kg an ICD system was implanted using epicardial sense-pacing leads and a superior vena caval lead as a subcutaneous defibrillator coil placed posterior on the left thorax. Shocks were delivered between the subcutaneous coil lead and the intraabdominally placed ICD can. This ICD array system has not been reported previously.
机译:先前健康的男婴在9周大时自发性心室颤动后得以复苏。超声心动图显示左心室三个肿瘤不适合完全切除。尽管使用抗心律不齐药物治疗,但室性心律失常仍在继续。当孩子4个月大,体重7公斤时,使用心外膜起搏导线和上腔静脉导线植入ICD系统,并将其作为皮下除颤器线圈置于左胸腔后。在皮下线圈导线和腹部放置的ICD罐之间传送了电击。此ICD阵列系统以前没有报告。