首页> 外文期刊>PACE: Pacing and clinical electrophysiology >Anomalous 'VVI' pacing of a dual chamber pacemaker when programmed in DDI(R) mode.
【24h】

Anomalous 'VVI' pacing of a dual chamber pacemaker when programmed in DDI(R) mode.

机译:在DDI(R)模式下编程时,双腔起搏器的“ VVI”起搏异常。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We describe the case of a dual chamber rate responsive pacemaker (Relay, model 294-03, Intermedics, Angleton, TX, USA) implanted in a 68-year-old male for sick sinus syndrome, which was not working properly when programmed in the DDIR mode, thus determining occasionally a sort of "VVI" pacing. However, the pacemaker performed well when programmed in the DDDR mode. We discovered that this was not a malfunction of a single device but rather a general behavior of this family of Intermedics dual chamber pacemakers (also not rate responsive), caused by a software problem.
机译:我们描述了一种双腔心率响应起搏器(继电器,型号294-03,Intermedics,美国德克萨斯州安格尔顿)的情况,该患者植入了68岁的男性患病性鼻窦综合症,当在DDIR模式,因此偶尔会确定一种“ VVI”起搏。但是,在DDDR模式下进行编程时,起搏器表现良好。我们发现这不是单个设备的故障,而是该Intermedics系列双腔起搏器的一般行为(也没有速率响应),这是由软件问题引起的。