PA/Lateral Chest X-Ray is Equivalent to Cine-Fluoroscopy for the Detection of Conductor Externalization in Defibrillation Leads

机译:PA /侧面胸部X射线等效于电影X线透视法,可检测除颤导线中的导体外化

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

BackgroundRiata and Riata ST defibrillation leads (St. Jude Medical, Sylmar, CA, USA) are susceptible to insulation defects with conductor externalization. Cine-fluoroscopy is considered to be the gold standard for the documentation of insulation defects, but similar detection rates have been reported for posterior-anterior (PA)/lateral chest x-ray (CXR) with zooming.
机译:背景Riia和Riata ST除颤导线(美国加利福尼亚州希尔玛市的St. Jude Medical公司)容易受到导体外在化的绝缘缺陷的影响。荧光透视法被认为是记录绝缘缺陷的金标准,但是对于后部放大(PA)/侧面胸部X射线(CXR)放大显示了相似的检测率。