'Tuned' defibrillation waveforms outperform 50/50% tilt defibrillation waveforms: a randomized multi-center study.

机译:“调谐”除颤波形优于50/50%倾斜除颤波形:一项随机的多中心研究。

获取原文
获取原文并翻译 | 示例

摘要

Introduction: A superior performance of a tuned waveform based on duration using an assumed cardiac membrane time constant of 3.5 ms and of a 50/50% tilt waveform over a standard 65/65% tilt waveform has been documented before. However, there has been no direct comparison of the tuned versus the 50/50% tilt waveforms. Methods: In 34 patients, defibrillation thresholds (DFTs) for tuned versus 50/50% tilt waveforms in a random order were measured by using the optimized binary search method. High voltage lead impedance was measured and used to select the pulse widths for tuned and 50/50% tilt defibrillation waveforms. Results: Delivered energy (7.3 +/- 4.6 J vs 8.7 +/- 5.3 J, P = 0.01), stored energy (8.2 +/- 5.1 J vs 9.7 +/- 5.6 J, P = 0.01), and delivered voltage (405.9 +/- 121.7 V vs 445.0 +/- 122.6 V, P = 0.008) were significantly lower for the tuned than for the 50/50% tilt waveform. In four patients with DFT >/=15 J, the tuned waveform lowered the mean energy DFT by 2.8 J and mean voltage DFT by 45 V. For all patients, the mean peak delivered energy DFT was reduced from 29 J to 22 J (24% decrease). Multiple regression analysis showed that a left ventricular ejection fraction <20% is a significant predictor of this advantage. Conclusion: Energy and voltage DFTs are lowered with an implantable cardioverter defibrillator that uses a tuned waveform compared to a standard 50% tilt biphasic waveform. (PACE 2007; 30:S139-S142).
机译:简介:以前已经证明过,基于假设的3.5 ms心脏膜时间常数的基于持续时间的调谐波形和优于标准65/65%倾斜波形的50/50%倾斜波形的优越性能。但是,尚没有直接比较调谐波形和50/50%倾斜波形的方法。方法:在34例患者中,使用优化的二分法搜索方法测量了相对于50/50%倾斜波形的随机去纤颤阈值(DFT)。测量了高压引线阻抗,并将其用于选择调谐和50/50%倾斜除颤波形的脉冲宽度。结果:输送的能量(7.3 +/- 4.6 J vs 8.7 +/- 5.3 J,P = 0.01),存储的能量(8.2 +/- 5.1 J vs 9.7 +/- 5.6 J,P = 0.01)和输送的电压(与50/50%倾斜波形相比,调谐后的405.9 +/- 121.7 V与445.0 +/- 122.6 V,P = 0.008)明显更低。在四名DFT> / = 15 J的患者中,调谐波形将平均能量DFT降低了2.8 J,将平均电压DFT降低了45V。对于所有患者,平均峰值递送能量DFT从29 J降低到22 J(24减少%)。多元回归分析表明,左心室射血分数<20%是此优势的重要预测指标。结论:与标准的50%倾斜双相波形相比,使用调谐波形的可植入心脏复律除颤器降低了能量和电压DFT。 (PACE 2007; 30:S139-S142)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号