首页> 外文期刊>PACE: Pacing and clinical electrophysiology >Inappropriate therapy from a defibrillator complicating transcoronary ablation of septal hypertrophy in a patient with hypertrophic obstructive cardiomyopathy.
【24h】

Inappropriate therapy from a defibrillator complicating transcoronary ablation of septal hypertrophy in a patient with hypertrophic obstructive cardiomyopathy.

机译:在患有肥厚性梗阻性心肌病的患者中,除纤颤器的不当治疗会使冠状动脉的间隔肥厚消融变得复杂。

获取原文
获取原文并翻译 | 示例
           

摘要

This case report describes a patient with obstructive hypertrophic cardiomyopathy who received therapy inappropriately from his implanted defibrillator, subsequent to transcoronary alcohol ablation for septal hypertrophy (TASH). Widening of the intracardiac electrogram postablation resulted in "double counting" of the intrinsic ventricular electrogram by the device and inappropriate tachycardia detection.
机译:该病例报告描述了患有梗阻性肥厚型心肌病的患者,该患者在行冠状动脉酒精性间隔肥厚(TASH)消融后,从植入的除颤器中接受了不合适的治疗。心内电图后消融的扩大导致该设备对固有心室电图的“重复计数”和不适当的心动过速检测。