To design practical planar high-T_c SQUID gradiometers for magnetocardiographs, a signal-to-noise (S/N) chart that will enable the gradiometers to detect P-waves was developed. The signal was estimated as the gradient field generated by the mean current dipole of P-waves of 20 healthy volunteers. The current dipoles were obtained by using a conventional magnetocardiograph comprising low-T_c SQUID gradiometers. On the other hand, the noise was estimated from studies on previously reported gradiometers based on the limit of intrinsic gradient field noise. The obtained S/N chart indicates that a gradiometer with a baseline longer than 12 mm is necessary for detecting P-waves without averaging. The reliability of the S/N chart was confirmed by detecting actual P-waves with 100-beat averaging by using a fabricated planar high-T_c SQUID gradiometer with a baseline of 6.75 mm.
机译:为了设计用于心电图仪的实用的平面高T_c SQUID梯度计,开发了使该梯度计能够检测P波的信噪比(S / N)图。该信号被估计为由20位健康志愿者的P波平均电流偶极子产生的梯度场。通过使用包括低T_c SQUID梯度仪的常规心电图仪获得电流偶极子。另一方面,根据固有梯度场噪声的极限,根据先前报告的梯度仪的研究估计了噪声。所获得的信噪比图表明,基线大于12 mm的梯度仪对于检测P波而不进行平均是必要的。通过使用基线为6.75 mm的平面高T_c SQUID梯度计制造的100拍均值检测实际P波,可以确认S / N图的可靠性。