首页> 外文期刊>Spectrochimica Acta, Part B. Atomic Spectroscopy >Limits of detection and decision. Part 4
【24h】

Limits of detection and decision. Part 4

机译:检测和决策的局限性。第4部分

获取原文
获取原文并翻译 | 示例

摘要

Probability density functions (PDFs) have been derived for a number of commonly used limit of detection definitions, including several variants of the Relative Standard Deviation of the Background-Background Equivalent Concentration (RSDB-BEC) method, for a simple linear chemical measurement system (CMS) having homoscedastic, Gaussian measurement noise and using ordinary least squares (OLS) processing. All of these detection limit definitions serve as both decision and detection limits, thereby implicitly resulting in 50% rates of Type 2 errors. It has been demonstrated that these are closely related to Currie decision limits, if the coverage factor, k, is properly defined, and that all of the PDFs are scaled reciprocals of noncentral t variates. All of the detection limits have well-defined upper and lower limits, thereby resulting in finite moments and confidence limits, and the problem of estimating the noncentrality parameter has been addressed. As in Parts 1-3, extensive Monte Carlo simulations were performed and all the simulation results were found to be in excellent agreement with the derived theoretical expressions. Specific recommendations for harmonization of detection limit methodology have also been made. (C) 2007 Elsevier B.V. All rights reserved.
机译:对于简单的线性化学测量系统,已经为许多常用的检测限定义导出了概率密度函数(PDF),包括背景-背景等效浓度相对标准偏差(RSDB-BEC)方法的几种变体( CMS)具有等高斯测量噪声,并使用普通最小二乘(OLS)处理。所有这些检测极限定义既可以作为决策极限,也可以作为检测极限,从而隐式地导致50%的2型错误率。已经证明,如果覆盖因子k得到正确定义,则这些与Currie决策限制密切相关,并且所有PDF都是非中心t变量的比例倒数。所有检测极限均具有明确定义的上限和下限,从而导致有限的力矩和置信度极限,并且解决了估计非中心性参数的问题。与第1-3部分一样,进行了广泛的蒙特卡洛模拟,发现所有模拟结果都与推导的理论表达式非常吻合。还提出了统一检测限方法的具体建议。 (C)2007 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号