首页> 外文期刊>Spectrochimica Acta, Part B. Atomic Spectroscopy >A multivariate study of the acid effect and the selection of internal standards for inductively coupled plasma mass spectrometry
【24h】

A multivariate study of the acid effect and the selection of internal standards for inductively coupled plasma mass spectrometry

机译:酸效应的多变量研究和电感耦合等离子体质谱法内标的选择

获取原文
获取原文并翻译 | 示例
           

摘要

The effects of nitric acid concentration and liquid flow rate on the signals obtained with an inductively coupled plasma mass spectrometer, operated with a microconcentric nebulizer (MCN-100), have been studied by the use of partial least squares regression. Thirteen elements, covering the mass range 51-238, were studied. The nitric acid concentration was varied in the range 0.14-2.8 mol l~(-1) and the acid effect was found to vary significantly between the elements - depending mainly on the value of the first ionisation potential of the element. The effect of the liquid flow rate was studied within the range 34.5-172.5 #mu#l min~(-1) and it was only significant (P = 0.05) for ~(107)Ag and ~(109)Ag. It is possible to compensate completely for the acid matrix effect by the use of a proper internal standard. The use of multivariate analysis methods. e.g. partial least squares regression is recommended in order to choose the most suitable internal standards.
机译:已经通过使用偏最小二乘回归研究了硝酸浓度和液体流速对使用微同心雾化器(MCN-100)操作的电感耦合等离子体质谱仪获得的信号的影响。研究了覆盖质量范围51-238的13个元素。硝酸浓度在0.14-2.8 mol l〜(-1)范围内变化,并且发现酸效应在元素之间显着变化-主要取决于元素的第一电离电势值。在34.5-172.5#mu#l min〜(-1)范围内研究了液体流速的影响,对于〜(107)Ag和〜(109)Ag仅显着(P = 0.05)。通过使用适当的内标可以完全补偿酸基质的影响。使用多元分析方法。例如为了选择最合适的内标,建议采用偏最小二乘回归。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号