Background correction errors for lead in the presence of phosphate with Zeeman graphite furnace atomic absorption spectrometry

机译:用塞曼石墨炉原子吸收光谱法对存在磷酸盐的铅的背景校正误差

获取原文
获取原文并翻译 | 示例
           

摘要

Background overcorrection problems with lead in the presence of phosphate in Zeeman graphite furnace atomic absorption spectrometry are investigated. Results show that the overcorrection is not affected by changing the atomization temperature in the furnace program, but is significantly reduced by decreasing the strength of the magnetic field. This observation further supports the assumption that the background overcorrection is caused by the presence of a molecular species (such as PO) formed during the heating process having molecular absorption bands which are subject to Zeeman-effect splitting. The overcorrection error can be diminished by deceasing the amount of phosphate or by replacing the standard transverse-heated graphite atomizer (THGA) tube with an end-capped tube. An accurate method for the determination of Pb in bone using a transverse heated furnace (a P-E 4100ZL) is described. (C) 1998 Elsevier Science B.V. All rights reserved. [References: 14]
机译:研究了塞曼石墨炉原子吸收光谱法中存在磷酸盐时铅的背景超校正问题。结果表明,过校正不受炉程序中雾化温度的改变的影响,但通过降低磁场强度可以大大降低。该观察结果进一步支持了这样的假设,即背景超校正是由于在加热过程中形成的分子物质(例如PO)的存在而引起的,该分子物质具有受塞曼效应分裂的分子吸收带。可以通过减少磷酸盐的量或通过将标准的横向加热石墨雾化器(THGA)管替换为带端盖的管来减少过校正误差。描述了使用横向加热炉(P-E 4100ZL)测定骨中Pb的准确方法。 (C)1998 Elsevier Science B.V.保留所有权利。 [参考:14]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号