摘要

We read with interest the recent paper by Bjartmarz and Rehncrona, 'Comparison of accuracy and precision between frame-based and frameless stereotactic navigation for deep brain stimulation electrode implantation' . We applaud the authors for endeavoring to provide a direct comparison of the two techniques; however, their methodology is missing certain details which we have found helpful in achieving optimal accuracy during frameless DBS placement.We believe it is vitally important to take careful note of registration accuracy and predicted volume of 1 mm error. There is no mention of these figures in the present study, making it difficult to compare this common source of error between studies.
机译:我们感兴趣地阅读了Bjartmarz和Rehncrona的最新论文,``用于深部脑刺激电极植入的基于帧和无帧立体定位导航的准确性和精确度的比较''。我们赞扬作者努力提供两种技术的直接比较;但是,他们的方法缺少某些细节,我们发现这些细节有助于在无框DBS放置期间实现最佳精度。我们认为,务必特别注意套准精度和预计的1 mm误差量,这一点至关重要。在本研究中没有提及这些数字,因此很难在研究之间比较这种常见的误差来源。