Reproducibility of measuring the shape and three-dimensional position of cervical vertebrae in upright position using the EOS stereoradiography system.

机译:使用EOS立体放射成像系统测量直立位置的颈椎骨的形状和三维位置的可重复性。

获取原文
获取原文并翻译 | 示例
           

摘要

STUDY DESIGN: An interobserver and intraobserver reproducibility study of the use of EOS stereoradiography system at the cervical spine. OBJECTIVE: To investigate reproducibility of the determination of the vertebral shape, position, and orientation of C3-C7 vertebrae in vivo using the EOS stereoradiography system. SUMMARY OF BACKGROUND DATA: Since CT and MRI 3-dimensional (3D) analysis of the spinal architecture are done in supine position, measurements of the relative position of the cervical vertebrae in vivo in standing position requires stereoradiography. The innovative EOS system is an accurate and promising tool for stereoradiography. Its reproducibility at the cervical spine had to be assessed. METHODS: Twenty volunteers had biplanar radiographs of the cervical spine. C3-C7 were reconstructed by 2 observers, 2 times each. Each reconstruction was compared with the corresponding average object in term of shape (point-to-surface distance) and position (Tx, Ty, Tz, Rx, Ry, Rz in a local screw-frame). RESULTS: The 95% confidence interval of the error in shape was 1.83 mm. It was 0.84 mm, 1.42 mm, 0.58 mm, 2.53 degrees, 2.34 degrees, and 3.24 degrees for the position in Tx, Ty, Tz, Rx, Ry, and Rz. Intraobserver differences were not significant. Interobserver differences were significant for the shape and in Tx, Rx, and Ry (0.9 mm, 0.54 mm, 0.33 degrees, and 0.28 degrees). CONCLUSION: Overall reproducibility favorably compared with other imaging methods, whereas significant interobserver disagreements were narrow and partial.
机译:研究设计:颈椎间使用EOS立体放射成像系统的观察者间和观察者内可重复性研究。目的:探讨使用EOS立体放射成像系统测定体内C3-C7椎骨的椎体形状,位置和方向的可重复性。背景数据摘要:由于在仰卧位置进行了脊柱结构的CT和MRI 3维(3D)分析,因此,站立时体内颈椎相对位置的测量需要进行立体放射照相。创新的EOS系统是用于立体放射线照相术的准确且有希望的工具。必须评估其在颈椎的可重复性。方法:二十名志愿者对颈椎进行了双平面X线摄片。 C3-C7由2位观察员重建,每位2次。在形状(点到表面的距离)和位置(局部螺钉框架中的Tx,Ty,Tz,Rx,Ry,Rz)方面,将每个重建与相应的平均对象进行比较。结果:形状误差的95%置信区间为1.83 mm。 Tx,Ty,Tz,Rx,Ry和Rz中的位置分别为0.84 mm,1.42 mm,0.58 mm,2.53度,2.34度和3.24度。观察者间差异不显着。观察者之间的差异对于形状以及Tx,Rx和Ry均显着(0.9 mm,0.54 mm,0.33度和0.28度)。结论:与其他成像方法相比,总体可重复性良好,而观察者之间的重大分歧是狭窄和局部的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号