首页> 外文期刊>Spine >Three-dimensional rotational X-ray imaging for spine surgery: a quantitative validation study comparing reconstructed images with corresponding anatomical sections.
【24h】

Three-dimensional rotational X-ray imaging for spine surgery: a quantitative validation study comparing reconstructed images with corresponding anatomical sections.

机译:脊柱外科手术的三维旋转X射线成像:定量验证研究,将重建的图像与相应的解剖切片进行比较。

获取原文
获取原文并翻译 | 示例
           

摘要

STUDY DESIGN: A validation study was done in which reconstructed three-dimensional rotational x-ray images were quantitatively compared with corresponding anatomic sections. OBJECTIVES: To assess the accuracy of reconstructed images acquired on a three-dimensional rotational x-ray imaging device. SUMMARY OF BACKGROUND DATA: Minimally invasive procedures have proven quite successful as alternatives for a multitude of open treatments. An unfavorable property of this type of procedure is a lack of direct vision of the operating area. Three-dimensional rotational x-ray imaging may be able to merge the advantages of computed tomography and fluoroscopy: real-time two-dimensional projections for fast visual feedback and three-dimensional reconstructions for detailed volumetric imaging of complex anatomy. METHODS: Twenty traumatic thoracolumbar burst fractures were created and underwent pedicle screw instrumentation and balloon vertebroplasty. Subsequently, a three-dimensional dataset was obtained, and the midsagittal image was reconstructed. The specimens were sliced, and photographs were obtained. Multiple parameters on the reconstructed images and photographs were measured two times by two observers using a graphical method. The differences and standard deviations were calculated for the corresponding parameters and for the intraobserver data. RESULTS: The mean difference between the corresponding values ranged between -1.1 and 2.1 mm for all parameters. The standard deviation for the differences per parameter ranged between 1.2 and 3.2 mm. The intraobserver differences ranged from -0.8 to 1.4 mm, and the standard deviation varied between 0.4 and 2.4 mm. CONCLUSIONS: The reconstructed midsagittal images were accurate in all measured parameters. The three-dimensional rotational x-ray technique may prove to be valuable for less invasive spine surgery.
机译:研究设计:进行了一项验证研究,其中将重建的三维旋转X射线图像与相应的解剖切片进行了定量比较。目的:评估在三维旋转X射线成像设备上获取的重建图像的准确性。背景数据概述:微创手术已被证明是非常成功的多种开放治疗方法。这种程序的不利特性是缺乏对操作区域的直接观察。三维旋转X射线成像可能能够融合计算机断层扫描和荧光透视的优势:用于快速视觉反馈的实时二维投影和用于复杂解剖结构的详细体积成像的三维重建。方法:创建20例创伤性胸腰椎爆裂骨折,并进行椎弓根螺钉器械植入和气囊椎体成形术。随后,获得三维数据集,并重建矢状正中图像。切片标本,并获得照片。两名观察员使用图形方法对重建的图像和照片上的多个参数进行了两次测量。计算相应参数和观察者内部数据的差异和标准偏差。结果:所有参数的相应值之间的平均差在-1.1和2.1 mm之间。每个参数差异的标准偏差范围为1.2到3.2 mm。观察者内差异在-0.8至1.4 mm之间,标准偏差在0.4至2.4 mm之间变化。结论:重建的矢状位中影在所有测量参数中均是准确的。三维旋转X射线技术可能证明对侵入性较小的脊柱外科手术很有价值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号