Noninvasive carbon dioxide monitoring during neurosurgical procedures in adults: end-tidal versus transcutaneous techniques.

机译:成人神经外科手术过程中的无创二氧化碳监测:潮气末与经皮技术。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: We prospectively compared transcutaneous (TC) versus end-tidal (ET) carbon dioxide monitoring during neurosurgical procedures in adults. METHODS: After calibration and an equilibration time for the TC-CO2 monitor, arterial blood gas (ABG) values were obtained as clinically indicated. The PaCO2 values were compared with the values recorded by the noninvasive monitors (TC and ET). RESULTS: The ET-CO2 to PaCO2 difference was 6.1 +/- 5.6 mm Hg, and the TC-CO2 to PaCO2 difference was 3.7 +/- 2.9 mm Hg. The difference between the PaCO2 and ET-CO2 was 3 mm Hg or less in 17 of 57 values, while the difference between the PaCO2 and TC-CO2 was 3 mm Hg or less in 35 of 57 values. Linear regression analysis of ET-CO2 versus PaCO2 revealed a slope of 0.381 +/- 0.007. Linear regression analysis of TC-CO2 versus PaCO2 revealed a slope of 1.17 +/- 0.008. CONCLUSION: Transcutaneous CO2 monitoring provides a more accurate estimate of PaCO2 than ET-CO2 monitoring during neurosurgical procedures.
机译:背景:我们前瞻性地比较了成人神经外科手术过程中经皮(TC)与潮气(ET)二氧化碳的监测。方法:在校准并平衡了TC-CO2监测器的时间后,按照临床指示获得了动脉血气(ABG)值。将PaCO2值与无创监护仪(TC和ET)记录的值进行比较。结果:ET-CO2与PaCO2的差为6.1 +/- 5.6 mm Hg,而TC-CO2与PaCO2的差为3.7 +/- 2.9 mm Hg。 PaCO2和ET-CO2之间的差值为57个值中的17个,为3 mm Hg或更少,而PaCO2和TC-CO2之间的差值为57个值中的35个3 mm Hg或以下。 ET-CO2与PaCO2的线性回归分析显示斜率为0.381 +/- 0.007。 TC-CO2与PaCO2的线性回归分析显示斜率为1.17 +/- 0.008。结论:在神经外科手术过程中,经皮CO2监测比ET-CO2监测更准确地估算PaCO2。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号