Constrained testing conditions affect the axial rotation response of lumbar functional spinal units.

机译:受限的测试条件会影响腰椎功能性脊柱单位的轴向旋转响应。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

STUDY DESIGN: Human cadaveric spine specimens were tested in axial rotation using constrained and unconstrained methods. OBJECTIVES: To determine the degree to which constrained methods affect the response of the functional spinal unit in axial rotation at lumbar and lumbosacral levels. SUMMARY OF BACKGROUND DATA: A substantial controversy exists in the literature regarding the appropriateness of different testing methods. No study has been found in which the effect of constraint on axial rotation behavior was objectively examined. METHODS: Ten human cadaveric spine specimens (five L3-L4, five L5-S1) were tested in axial rotation, using both constrained and unconstrained methods. In the unconstrained test, pure moments were applied to the upper vertebra, and its complete three-dimensional motion was measured using an optoelectronic camera system. In the constrained test, the specimens were loaded in a fixed-axis servohydraulic test machine individually around five rotational axis positions within the vertebral body, and the rotational motion was measured. RESULTS: The rotational angles in the constrained tests were not different among the five rotational axis positions. However, the maximum rotation from the five axis positions was approximately 40% greater than the minimum rotation, a significant difference. The axial rotational motion of the unconstrained tests was always less than the maximum rotation measured in the constrained test. However, the total rotational angle using the helical axis of motion was not significantly different from the constrained angles. CONCLUSIONS: The large differences between maximum and minimum rotation angles demonstrate that the behavior of the functional spinal unit in axial rotation is sensitive to the axis's position but the location of the axis is not repeatable. This supports the use of unconstrained methods in spinal testing.
机译:研究设计:使用约束和非约束方法对人体尸体脊柱标本进行轴向旋转测试。目的:确定受限方法在腰椎和腰水平轴向旋转时影响功能性脊柱单位反应的程度。背景技术概述:关于不同测试方法的适当性,文献中存在着很大的争议。尚未发现客观研究约束对轴向旋转行为的影响的研究。方法:使用约束和非约束方法对十个人体尸体脊柱标本(五个L3-L4,五个L5-S1)进行了轴向旋转测试。在无约束测试中,将纯力矩施加到上椎骨,并使用光电照相机系统测量其完整的三维运动。在约束试验中,将试样分别围绕椎体内的五个旋转轴位置分别装在固定轴伺服液压试验机中,并测量旋转运动。结果:约束测试中的旋转角度在五个旋转轴位置之间没有差异。但是,从五个轴位置开始的最大旋转比最小旋转大约大40%,差异很大。无约束试验的轴向旋转运动总是小于在有约束试验中测得的最大旋转。但是,使用螺旋运动轴的总旋转角度与约束角度没有显着差异。结论:最大和最小旋转角度之间的巨大差异表明,功能性脊柱单位在轴向旋转中的行为对轴的位置敏感,但轴的位置不可重复。这支持在脊柱测试中使用不受限制的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号