Measurements of bone mineral density in the lumbar spine and proximal femur using lunar prodigy and the new pencil-beam dual-energy X-ray absorptiometry.

机译:使用月球神器和新型笔形束双能X射线吸收仪测量腰椎和股骨近端的骨矿物质密度。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVE: We evaluated the correlation of the absolute bone mineral density (BMD) values of the lumbar spine and standard sites of the proximal femur obtained from a Lunar Prodigy and the newly developed pencil-beam dual-energy X-ray absorptiometry (Dexxum). MATERIALS AND METHODS: Between June 2008 and December 2008, 79 Korean volunteers were enrolled. Measurements were obtained on the same day using both densitometers. The absolute BMD values (g/cm(2)) from the two densitometers were evaluated using Pearson's correlation analysis with Bonferroni's correction for the three clinically important sites. In order to evaluate precision, we performed duplicate Dexxum measurements, and calculated the within-subject coefficient of variation (WSCV). RESULTS: The Pearson's correlation coefficient (r) of BMD values for the total proximal femur, femoral neck, and lumbar spine by the two densitometers were 0.926, 0.948, and 0.955 respectively, and the null hypotheses of r = 0.8 were all rejected (p < 0.001 by one-sided Z-test with Fisher's z-transformation for each site). The T-scores (r >== 0.842) and Z-scores (r >== 0.709) also showed strong positive correlations. The duplicate BMD values of Dexxum showed a high level of precision (WSCV <== 4.27%). CONCLUSION: Dexxum measurements of BMD, T-scores, and Z-scores showed a strong linear correlation with those measured on Lunar Prodigy.
机译:目的:我们评估了从Lunar Prodigy和新开发的笔束双能X射线吸收仪(Dexxum)获得的腰椎和股骨近端标准部位的绝对密度,以及腰椎的绝对矿物质密度(BMD)值的相关性。材料与方法:2008年6月至2008年12月,招募了79名韩国志愿者。使用两个光密度计在同一天获得测量值。使用皮尔逊相关分析和Bonferroni校正对三个临床上重要的部位,评估了两个光密度计的绝对BMD值(g / cm(2))。为了评估精度,我们进行了重复的Dexxum测量,并计算了受试者内部变异系数(WSCV)。结果:两个密度计的总股骨近端,股骨颈和腰椎的BMD值的皮尔逊相关系数(r)分别为0.926、0.948和0.955,并且r = 0.8的零假设均被拒绝(p通过单侧Z检验(每个站点的Fisher的z变换)<0.001)。 T分数(r> == 0.842)和Z分数(r> == 0.709)也显示出强正相关。 Dexxum的重复BMD值显示出很高的精度(WSCV <== 4.27%)。结论:BMD,T分数和Z分数的Dexxum测量值与Lunar Prodigy上的测量值显示出很强的线性相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号