首页> 外文期刊>Central European Journal of Chemistry >Evaluation of various sample pre-treatment methods for total and inorganic mercury determination in biological certified reference materials by CVAAS technique
【24h】

Evaluation of various sample pre-treatment methods for total and inorganic mercury determination in biological certified reference materials by CVAAS technique

机译:利用CVAAS技术评估生物认证参比物质中总汞和无机汞的各种样品预处理方法的评估

获取原文
获取原文并翻译 | 示例
           

摘要

Sample preparation methods for non-separation cold vapor atomic absorption spectrometry (CVAAS) sequential inorganic mercury speciation in biological certified reference materials (CRMs) were investigated. The methylmercury concentration was calculated as the difference between total and inorganic mercury. Microwave-assisted decomposition method, and three ultrasonic extraction pro?cedures based on acid leaching with HCl and HCOOH and solubilization with TMAH were employed as sample preparation methods. The replacement of a sample decomposition procedure by extraction prior to analysis by CVAAS, as well as the aspect of speciation analysis is discussed. The limits of detection in the sample were determined as 50 and 10 ng L~(-1) for inorganic and total mercury, which corresponds to absolute detection limits of 40 and 8ng g~(-1) for inorganic and total mercury, respectively. The results were in good agreement with the 95% confidence level t-test of the certified values for total and inorganic mercury in the reference materials investigated. From the analysis of the CRMs, it was evident that the difference between the total and inorganic mercury concentrations agrees with the methylmercury concentration. The relative standard deviation was better than 11% for most of the samples.
机译:研究了在生物认证参考材料(CRM)中进行非分离式冷蒸气原子吸收光谱法(CVAAS)连续无机汞形态形成的样品制备方法。甲基汞浓度计算为总汞和无机汞之差。微波辅助分解法和基于HCl和HCOOH的酸浸并以TMAH增溶为基础的三种超声提取程序被用作样品制备方法。讨论了在通过CVAAS分析之前通过提取来代替样品分解程序以及形态分析的方面。样品中无机汞和总汞的检出限分别确定为50和10 ng L〜(-1),分别对应于无机汞和总汞的绝对检出限分别为40和8ng g〜(-1)。结果与所研究参考物质中总汞和无机汞的认证值的95%置信水平t检验吻合良好。从CRM的分析可以看出,总汞浓度和无机汞浓度之间的差异与甲基汞浓度一致。大多数样品的相对标准偏差均优于11%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号