A PRESSURE TRANSDUCER USING A METALLIC TRIPLE-BEAM TUNING FORK

机译:使用金属三束调谐叉的压力传感器

获取原文
获取原文并翻译 | 示例
           

摘要

A pressure transducer has been developed using a metallic resonating tuning fork. The fork has ken mounted horizontally across two pillars which are an integral part of the diaphragm of a pressure transducer. The displacement of the. diaphragm induces an axial force in the fork. The applied stress causes the natural frequency of the fork to increase, The experimental pressure transducer has a linear characteristic over its 6 bar range with repeatability better than 0.1%, and hysteresis of 0.1% of the range. [References: 7]
机译:已经使用金属谐振音叉开发了压力传感器。叉子在水平方向上跨过两个支柱安装,这两个支柱是压力传感器膜片的组成部分。的位移。隔膜在货叉中产生轴向力。施加的应力会导致货叉的固有频率增加。实验压力传感器在其6 bar范围内具有线性特性,重复性优于0.1%,滞后值为该范围的0.1%。 [参考:7]