首页> 外文期刊>Sensor Letters: A Journal Dedicated to all Aspects of Sensors in Science, Engineering, and Medicine >Molecularly Imprinted Polymers (MIP) Based Electrochemical Sensor for Detection of Urea and Creatinine
【24h】

Molecularly Imprinted Polymers (MIP) Based Electrochemical Sensor for Detection of Urea and Creatinine

机译:基于分子印迹聚合物(MIP)的电化学传感器,用于检测尿素和肌酐

获取原文
获取原文并翻译 | 示例
           

摘要

In this work, sensors based on molecularly imprinted polymers (MIP) for creatinine and urea selective determination using solvent evaporation processing of poly(ethylene-co-vinyl alcohol) (EVAL) to form molecularly imprinted polymer s (MIP) were fabricated. The chemical behaviour of the thin film (MIP) was characterized using FTIR spectrophotometry. The carbonyl functions assigned on the spectra confirmed capture and removal (after rinsing with ethanol) of template molecule. Electrochemical impedance method was employed as transduction technique. The EIS (Electrochemical Impedance Spectroscopy) sensor response shows a limit of detection of 10 ng/mL with linear range from 0.02μg/mL to 3 μg/mL for urea and 40 ng/mL of detection limit with linear range from 0.05 μg/mL to 2 μg/mL for creatinine with a good specificity of detection.
机译:在这项工作中,制造了基于分子印迹聚合物(MIP)的肌酐和尿素选择性测定传感器,该溶剂使用聚(乙烯-共-乙烯醇)(EVAL)的溶剂蒸发工艺形成了分子印迹聚合物(MIP)。使用FTIR分光光度法对薄膜(MIP)的化学行为进行了表征。光谱上分配的羰基官能团证实了模板分子的捕获和去除(用乙醇冲洗后)。电化学阻抗法被用作转导技术。 EIS(电化学阻抗谱)传感器响应显示尿素线性范围从0.02μg/ mL到3μg/ mL的检测限为10 ng / mL,线性范围从0.05μg/ mL的检测限为40 ng / mL血肌酐浓度低至2μg/ mL,具有良好的检测特异性。