首页> 外文期刊>Scandinavian journal of clinical and laboratory investigation. >Determination of melatonin in saliva using automated solid-phase extraction, high-performance liquid chromatography and fluorescence detection.
【24h】

Determination of melatonin in saliva using automated solid-phase extraction, high-performance liquid chromatography and fluorescence detection.

机译:使用自动固相萃取,高效液相色谱和荧光检测法测定唾液中的褪黑素。

获取原文
获取原文并翻译 | 示例
           

摘要

A sensitive bioanalytical method for the determination of melatonin in saliva by solid-phase extraction (SPE), high-performance liquid chromatography (HPLC) and fluorescence detection has been developed and validated. Saliva was collected with a Salivette sampling device (Sarstedt) and a mixed-mode SPE column was used for the extraction of melatonin and internal standard (N-acetyl-6-methoxytryptamine) from the saliva. Chromatographic separation was performed using a HyPurity C18 LC column (150 x 2.1 mm) with mobile phase acetonitrile-ammonium hydrogen carbonate buffer, 0.015 M, pH 6.8 (23:77, v/v). Excitation and emission wavelengths were set to 285 nm and 345 nm, respectively. The within-day precision for the method at 50 pmol/L was 7.9 % and the between-day precision was 10.5 %. The limit of quantification was 50 pmol/L.
机译:建立并验证了一种灵敏的生物分析方法,可通过固相萃取(SPE),高效液相色谱(HPLC)和荧光检测法测定唾液中的褪黑激素。用Salivette采样装置(Sarstedt)收集唾液,并使用混合模式SPE柱从唾液中提取褪黑激素和内标(N-乙酰基-6-甲氧基色胺)。使用HyPurity C18 LC色谱柱(150 x 2.1 mm)和流动相乙腈-碳酸氢铵缓冲液,0.015 M,pH 6.8(23:77,v / v)进行色谱分离。激发和发射波长分别设置为285 nm和345 nm。该方法在50 pmol / L的日内精度为7.9%,日间精度为10.5%。定量限为50 pmol / L。