A Comparative Study of Audiologic Outcomes for Two Transcutaneous Bone-Anchored Hearing Devices

机译:两种经皮骨锚式听力设备的听力学结果的比较研究

获取原文
获取原文并翻译 | 示例
           

摘要

Objective:In 1977, Tjellstrom et al. used the concept of osseointegration and implanted patients with a titanium screw as part of the first bone-anchored hearing solution. Despite good hearing outcomes with the percutaneous device, there has been a drive for the development of transcutaneous bone conduction systems because of soft tissue complications, aesthetics, and loss or failure of the abutment. This study compares audiologic and quality of life questionnaire outcomes (author-designed questionnaire) for two transcutaneous bone conduction implants.Study Design:Cross-sectional cohort study.Setting:Tertiary referral hospital and national audiology service.Patients:Twelve patients, 10 children and two adults. Six recipients of each device.Main Outcome Measures:Pure-tone audiometry, aided thresholds, speech discrimination scores, and quality of life questionnaire results.Results:The unaided four-frequency average air conduction for affected ears was 60.8dB HL for the Attract group and 57.8 dBHL for the Sophono group; these improved to mean aided thresholds of 30.6dB HL and 29.8dB HL, respectively. Therewas no statistical difference between the speech discrimination scores for the two devices in quiet at 55dB SPL (p=0.33) orin noise (p=0.87).Conclusion:Both systems provide audiologic benefit compared with the unaided situation. Comparing small numbers of patients, there was no significant difference in aided thresholds or speech discrimination scores between these two transcutaneous bone-anchored implants. All Attract and Sophono users reported improvement in quality of life and would recommend their device to others in a similar situation.
机译:目的:1977年,Tjellstrom等人。使用骨整合的概念并将钛螺钉植入患者体内,这是第一个骨锚式听力解决方案的一部分。尽管使用经皮装置具有良好的听力效果,但由于软组织并发症,美观以及基台的丧失或衰竭,仍推动了经皮骨传导系统的发展。这项研究比较了两种经皮骨传导植入物的听力学和生活质量问卷结果(作者设计的问卷)。研究设计:横断面队列研究背景:三级转诊医院和国家听力学服务患者:12名患者,10名儿童和两个大人。每个设备有六个接收者。主要观察指标:纯音测听,辅助阈值,语音辨别力得分和生活质量调查问卷结果。结果:Attraction组的独立四频平均空气传导率为受影响的耳朵60.8dB HL Sophono组为57.8 dBHL;这些分别改善了平均辅助阈值30.6dB HL和29.8dB HL。在安静的55dB SPL(p = 0.33)或噪声(p = 0.87)时,这两种设备的语音辨别力得分之间没有统计学差异。比较少量患者,这两种经皮骨锚固植入物的辅助阈值或语音辨别力得分均无显着差异。所有Attract和Sophono用户均报告生活质量有所改善,并会向处于类似情况的其他人推荐其设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号