Computer-assisted audiometry versus manual audiometry.

机译:计算机辅助测听与手动测听。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE: The Otogram is an automated computer-assisted audiometer that allows patients to self-administer audiometry for their pure-tone audiogram. There has been no research published in a peer-reviewed journal, validating its use in an otology outpatient clinic. We therefore proposed to investigate and compare the inter-rater and intrarater accuracy and reliability of audiologists and of the Otogram in an English-speaking British population. DESIGN: Prospective nonrandomized validation study. SETTING: Secondary otolaryngology center and otology outpatient clinic. PARTICIPANTS: Forty-eight NHS patients referred to an otology outpatient clinic. MAIN OUTCOME MEASURES: Each patient had 2 pure-tone audiograms. Hearing thresholds in decibels hearing level were ascertained by fully trained British audiologists and by the Otogram. RESULTS: Using the weighted kappa statistic, the level of agreement in air-conduction (kappa = 0.965) and bone-conduction (kappa = 0.927) thresholds between the audiologist and the Otogram on the same patient was equivalent to the inter-rater level of agreement between pairs of audiologists. Approximately 94% of air-conduction thresholds and 91% of bone-conduction thresholds measured by the Otogram fell within 10 dB of thresholds measured by an audiologist. Intrarater comparisons between audiologists were very good for air-conduction (kappa = 0.978) and bone-conduction (kappa = 0.964). The intrarater level of agreement between repeated Otogram thresholds was just as good for air-conduction (kappa = 0.974) and bone-conduction (kappa = 0.945) thresholds. CONCLUSION: The Otogram is just as reliable as audiologists at determining hearing thresholds. We recommend that the Otogram can be safely used in a controlled clinical setting supervised by audiologists.
机译:目的:Otogram是一种自动的计算机辅助听力计,可让患者针对其纯音听力图进行自我测听。尚未在同行评审的期刊上发表过任何研究,证实了其在耳科门诊诊所中的使用。因此,我们建议调查和比较在英语为英语的英国人群中,听力学家和Otogram的评定者之间和评定者之间的准确性和可靠性。设计:前瞻性非随机验证研究。地点:二级耳鼻喉科中心和耳科门诊。参与者:48名NHS患者转诊至耳科门诊诊所。主要观察指标:每例患者均进行2次纯音听力图检查。经过充分训练的英国听力学家和Otogram确定了分贝听力水平的听力阈值。结果:使用加权kappa统计量,同一位患者的听觉医师和Otogram之间的空气传导(kappa = 0.965)和骨传导(kappa = 0.927)阈值的一致性水平等同于成对的听力学家之间的协议。耳声图测量的大约94%的空气传导阈值和91%的骨传导阈值落在听觉医师测量的阈值的10 dB之内。听力学家之间的intrarater比较在空气传导(kappa = 0.978)和骨传导(kappa = 0.964)方面非常好。重复的Otogram阈值之间的评估者内部一致性水平对于空气传导(kappa = 0.974)和骨传导(kappa = 0.945)阈值同样有效。结论:在确定听力阈值时,耳麦图谱与听力学家一样可靠。我们建议可以在听力学家的监督下,在受控的临床环境中安全使用Otogram。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号