Clinical evaluation of the Shin-Nippon NVision-K 5001/Grand Seiko WR-5100K autorefractor.

机译:Shin-Nippon NVision-K 5001 / Grand Seiko WR-5100K自动验光仪的临床评估。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: A clinical evaluation of the Shin-Nippon NVision-K 5001 (also branded as the Grand Seiko WR-5100K) autorefractor (Japan) was performed to examine validity and repeatability compared with subjective refraction and Javal-Schiotz keratometry. METHODS: Measurements of refractive error were performed on 198 eyes of 99 subjects (aged 23.2 +/- 7.4 years) subjectively (noncycloplegic) by one masked optometrist and objectively with the NVision-K autorefractor by a second optometrist. Keratometry measurements using the NVision-K were compared with the Javal-Schiotz keratometer. Intrasession repeatability of the NVision-K was also assessed on all 99 subjects together with intersession repeatability on a separate occasion separated by 7 to 14 days. RESULTS: Refractive error as measured by the NVision-K was found to be similar (p = 0.67) to subjective refraction (difference, 0.14 +/- 0.35 D). It was both accurate and repeatable over a wide prescription range (-8.25 to +7.25 D). Keratometry as measured by the NVision-K was found to be similar (p > 0.50) to the Javal-Schiotz technique in both the horizontal and vertical meridians (horizontal: difference, 0.02 +/- 0.09 mm; vertical: difference, 0.01 +/- 0.14 mm). There was minimal bias, and the results were repeatable (horizontal: intersession difference, 0.00 +/- 0.09 mm; vertical: intersession difference, -0.01 +/- 0.12 mm). CONCLUSION: The open-view arrangement of the Shin-Nippon NVision-K 5001 facilitates the measurement of static refractive error and the accommodative response to real-world stimuli. Coupled with its accuracy, repeatability, and capability to measure corneal curvature, it is a valuable addition to objective instrumentation currently available to the optometrist and researcher.
机译:目的:对Shin-Nippon NVision-K 5001(也称为Grand Seiko WR-5100K)自动验光仪(日本)进行了临床评估,以比较主观验光和Javal-Schiotz角膜曲率法的有效性和可重复性。方法:由一名蒙面验光师主观(非睫状肌麻痹)对99位受试者(23.2 +/- 7.4岁)的198只眼睛进行屈光不正的测量,由另一名验光师客观地使用NVision-K自动验光仪进行屈光不正的测量。将使用NVision-K进行的角速度测量与Javal-Schiotz角速度计进行了比较。还评估了所有99位受试者的NVision-K术中可重复性,以及在7至14天分开的单独情况下的术中可重复性。结果:NVision-K测量的屈光不正与主观屈光(相差0.14 +/- 0.35 D)相似(p = 0.67)。在很宽的处方范围(-8.25至+7.25 D)内,它既准确又可重复。通过NVision-K测量的角膜曲率在水平和垂直子午线上均与Javal-Schiotz技术相似(p> 0.50)(水平:差异0.02 +/- 0.09 mm;垂直:差异0.01 + / -0.14毫米)。偏差最小,结果可重复(水平:闭会间差异为0.00 +/- 0.09 mm;垂直:闭会间差异为-0.01 +/- 0.12 mm)。结论:Shin-Nippon NVision-K 5001的开视野布置便于测量静态屈光误差和对真实刺激的适应性反应。加上其准确性,可重复性和测量角膜曲率的能力,它是对验光师和研究人员当前可用的客观仪器的宝贵补充。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号