Reproducibility of tear meniscus measurement by Fourier-domain optical coherence tomography: a pilot study.

机译:傅里叶域光学相干断层扫描术测量泪液半月板的可重复性:一项初步研究。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND OBJECTIVE: To study the reproducibility of tear meniscus measurement with high-speed high-resolution Fourier-domain optical coherence tomography (FD-OCT). PATIENTS AND METHODS: Twenty normal participants were enrolled in this prospective study. The lower tear meniscus in the right eye of each subject was imaged by vertical scans centered on the inferior cornea and the lower eyelid using an FD-OCT system (RTVue; Optovue, Inc., Fremont, CA) with a corneal adaptor. The system performs 26,000 axial scans per second and has a 5-micron axial resolution. Each subject was examined at two visits 30 to 60 days apart. Each eye was scanned twice on each visit. The scans were taken 2 seconds after a blink. The lower meniscus height, depth, and cornea-meniscus angle were measured with a computer caliper. The cross-sectional area was calculated using a two-triangle approximation. RESULTS: The between-visits coefficient of variation was 17.5%, 18.0%, 35.5%, and 12.2% for meniscus height, depth, area, and angle, respectively. The intraclass correlations for these parameters were 0.605, 0.558, 0.567, and 0.367, respectively. CONCLUSION: FD-OCT measures lower tear meniscus dimensions and area with higher between-visits reproducibility than previous OCT instruments. FD-OCT may be a useful way to measure dry eye severity and treatment effectiveness.
机译:背景与目的:研究高速高分辨率傅里叶域光学相干断层扫描(FD-OCT)测量泪液弯月面的可重复性。患者与方法:20名正常参与者参加了这项前瞻性研究。使用带有角膜适配器的FD-OCT系统(RTVue; Optovue,Inc.,Fremont,CA),通过以下角膜和下眼睑为中心的垂直扫描,对每个受试者右眼的下泪液弯液面成像。该系统每秒执行26,000次轴向扫描,轴向分辨率为5微米。每个受试者在相隔30至60天的两次访问中接受检查。每次访问对每只眼睛扫描两次。眨眼后2秒钟进行扫描。使用计算机卡尺测量下半月板的高度,深度和角膜半月板角度。截面积使用两三角形近似计算。结果:半月板高度,深度,面积和角度的两次就诊变异系数分别为17.5%,18.0%,35.5%和12.2%。这些参数的类内相关性分别为0.605、0.558、0.567和0.367。结论:FD-OCT测量的泪液弯液面尺寸和面积比以前的OCT仪器低,两次访问之间的再现性更高。 FD-OCT可能是衡量干眼症严重程度和治疗效果的有用方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号