Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes.

机译:在健康的眼睛中,新型ICare回弹眼压计与其他便携式眼压计的准确性。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: The ICare (Tiolat Oy, Helsinki, Finland) is a new portable tonometer that measures intraocular pressure (IOP) with a new rebound method, in which a very light probe is used to make momentary contact with the cornea in slow motion. The purpose of this study is to assess the accuracy of the ICare IOP measurements by comparing them against other portable tonometers: Perkins applanation tonometer and Tono-Pen XL digital tonometer (Medtronic Solan, Jacksonville, FL). METHODS: Sixty-five young subjects were assessed with each of the tonometers. ICare tonometry was performed first, followed by Perkins applanation tonometry and Tono-Pen XL in a random order. Regression analysis was used to evaluate the relationship between the Perkins tonometer and the remaining tonometers used in this study. Tonometers were also compared by plotting the difference between the methods against the mean. The hypothesis of zero bias was examined by a paired t test and the 95% limits of agreement (LoA) were calculated. RESULTS: ICare and Tono-Pen XL significantly overestimate IOP when compared with Perkins applanation tonometry. The mean of the difference between Perkins and ICare and Perkins and Tono-Pen XL was (mean +/- standard deviation) -3.35 +/- 2.28 mm Hg and -2.78 +/- 2.53 mm Hg, respectively. The 95% LoA between Perkins tonometry and ICare tonometry were between -7.81 and +1.12 and between Perkins tonometry and Tono-Pen XL tonometry between -7.74 and +2.18. CONCLUSIONS: Compared with Perkins tonometry, the ICare tonometer allows clinicians to estimate IOP with a portable, rapid, and noninvasive method with similar reliability to that offered by Tono-Pen XL. Clinicians should be aware of the systematic overestimation of IOP with the ICare. Further research is needed to evaluate the performance of rebound tonometry in populations with higher IOP and assess the reliability of this technique in the early detection and follow up of glaucomatous patients.
机译:用途:ICare(芬兰赫尔辛基的蒂奥拉特·奥伊公司)是一种新型便携式眼压计,它采用一种新的回弹方法来测量眼内压(IOP),其中使用了非常轻的探针以慢速瞬时接触角膜。本研究的目的是通过与其他便携式眼压计(Perkins压平眼压计和Tono-Pen XL数字眼压计(Medtronic Solan,杰克逊维尔,佛罗里达州))进行比较来评估ICare IOP测量的准确性。方法:用每个眼压计评估了65名年轻受试者。首先进行ICare眼压测定,然后以随机顺序进行Perkins压平眼压测定和Tono-Pen XL。回归分析用于评估珀金斯眼压计与本研究中使用的其余眼压计之间的关系。还通过将方法之间的差异与平均值作图来比较眼压计。零偏的假设通过配对t检验进行了检验,并计算了95%的一致性极限(LoA)。结果:与Perkins压平眼压计相比,ICare和Tono-Pen XL大大高估了IOP。 Perkins和ICare以及Perkins和Tono-Pen XL之间的均值分别为(平均值+/-标准偏差)-3.35 +/- 2.28 mm Hg和-2.78 +/- 2.53 mm Hg。 Perkins眼压和ICare眼压之间的95%LoA在-7.81和+1.12之间,Perkins眼压和Tono-Pen XL眼压之间在-7.74和+2.18之间。结论:与Perkins眼压计相比,ICare眼压计使临床医生能够以一种便携式,快速且无创的方法估算眼压,其可靠性与Tono-Pen XL所提供的相似。临床医生应该意识到ICare对IOP的系统性高估。需要进一步的研究来评估回声眼压计在高眼压人群中的表现,并评估该技术在青光眼患者的早期发现和随访中的可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号