Introduction: Uninterrupted chest compressions are a key factor in determining resuscitation success. Interruptions to chest compression are often associated with defibrillation, particularly the need to stand clear from the patient during defibrillation. It has been suggested that clinical examination gloves may provide adequate electrical resistance to enable safe hands-on defibrillation in order to minimise interruptions. We therefore examined whether commonly used nitrile clinical examination gloves provide adequate resistance to current flow to enable safe hands-on defibrillation. Methods: Clinical examination gloves (Kimberly Clark KC300 Sterling nitrile) worn by members of hospital cardiac arrest teams were collected immediately following termination of resuscitation. To determine the level of protection afforded by visually intact gloves, electrical resistance across the glove was measured by applying a DC voltage across the glove and measuring subsequent resistance. Results: Forty new unused gloves (control) were compared with 28 clinical (non-CPR) gloves and 128 clinical (CPR) gloves. One glove in each group had a visible tear and was excluded from analysis. Control gloves had a minimum resistance of 120. kΩ (median 190. kΩ) compared with 60. kΩ in clinical gloves (both CPR (median 140. kΩ) and non-CPR groups (median 160. kΩ)). Discussion: Nitrile clinical examination gloves do not provide adequate electrical insulation for the rescuer to safely undertake 'hands-on' defibrillation and when exposed to the physical forces of external chest compression, even greater resistive degradation occurs. Further work is required to identify gloves suitable for safe use for 'hands-on' defibrillation.
机译:简介:不间断的胸部按压是决定复苏成功的关键因素。胸部按压中断通常与除颤相关,尤其是在除颤过程中需要与患者保持距离。已经建议临床检查手套可以提供足够的电阻,以实现安全的手动除颤,以最大程度地减少干扰。因此,我们检查了常用的丁腈临床检查手套是否对电流具有足够的抵抗力,以实现安全的手动除颤。方法:复苏停止后,立即收集医院心脏骤停小组成员戴的临床检查手套(Kimberly Clark KC300 Sterling腈)。为了确定由视觉上完整的手套提供的保护水平,通过在手套上施加直流电压并测量随后的电阻来测量手套上的电阻。结果:将40只新的未使用手套(对照)与28支临床(非CPR)手套和128支临床(CPR)手套进行了比较。每组一只手套有可见的眼泪,因此被排除在分析之外。对照手套的最小电阻为120.kΩ(中值190.kΩ),而临床手套(CPR(中值140.kΩ)和非CPR组(中值160.kΩ))的最小电阻为60.kΩ。讨论:丁腈临床检查手套不能为救助人员提供足够的电绝缘,使其无法安全地进行“手动”除颤,并且在受到外部胸部按压的物理作用力时,甚至会出现更大的电阻衰减。需要进一步的工作来确定适合于“动手”除颤的安全使用的手套。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号