首页> 外文期刊>Resuscitation. >A new universal laryngoscope blade: a preliminary comparison with Macintosh laryngoscope blades.
【24h】

A new universal laryngoscope blade: a preliminary comparison with Macintosh laryngoscope blades.

机译:新的通用喉镜刀片:与Macintosh喉镜刀片的初步比较。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The Dorges universal laryngoscope blade has several features designed to facilitate tracheal intubation. The number of laryngoscope blades may be reduced from four to two, or even one, which indicate less space requirement and costs. This new universal laryngoscope blade, has a lower profile (height 15 vs. 22 mm) than a Macintosh laryngoscope blade size 3 and 4, which may facilitate manoeuvring of the laryngoscope in the mouth. In random order, 40 non-anaesthesia senior house officers used a Macintosh laryngoscope blade size 3 or 4 in an adult airway management trainer, a Macintosh laryngoscope blade size 2 in a paediatric airway management trainer, and the Dorges universal laryngoscope blade for both airway management trainers to perform orotracheal intubation. The number of intubation attempts and failures was counted. Participants reported the laryngoscopic view according to Cormack and Lehane. The time from touching the laryngoscope to the first adequate lung insufflation was measured, and subjective assessment regarding handling of both blades was recorded. Number of intubation failures, the laryngoscopic view according to Cormack and Lehane, and subjective assessment was comparable between groups. Orotracheal intubation of the adult airway management trainer with the Dorges universal laryngoscope blade took significantly less time compared to the Macintosh laryngoscope blades (14 (7-57) vs. 20 (8-43) s; P<0.001); all other intubating times were comparable. In conclusion, in this model, the Dorges universal laryngoscope blade was comparable to the Macintosh laryngoscope blades size 2-4, and may save time, cost and space.
机译:Dorges通用喉镜刀片具有旨在方便气管插管的多种功能。喉镜刀片的数量可以从四个减少到两个,甚至一个,这意味着更少的空间需求和成本。这种新的通用喉镜刀片的外形比3号和4号Macintosh喉镜的外形更小(高度15对22毫米),这可能有助于在口中操纵喉镜。按照随机顺序,有40位非麻醉高级内务人员在成人气道管理训练器中使用了3或4号的Macintosh喉镜刀片,在儿科气道管理训练器中使用了2号的Macintosh喉镜刀片,以及用于两种气道管理的Dorges通用喉镜刀片培训师进行口气管插管。计算插管尝试次数和失败次数。根据Cormack和Lehane的报道,参与者报告了喉镜的观点。测量从触摸喉镜到第一次适当的肺吹气的时间,并记录有关两个刀片操作的主观评估。两组之间的插管失败次数,喉镜检查次数以及主观评估均相当。与Macintosh喉镜刀片相比,使用Dorges通用喉镜刀片对成人气道管理教练员进行气管插管所花费的时间明显更少(14(7-57)s对20(8-43)s; P <0.001);所有其他插管时间都是可比的。总之,在此模型中,Dorges通用喉镜刀片可与2-4号Macintosh喉镜刀片媲美,并且可以节省时间,成本和空间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号