Purpose: To evaluate energy response and visibility for thin-film TLDs in mammography. Materials and methods: The energy responses of GR-200F thin-film TLDs were investigated at mammographic energies from 24 to 32kVp for the Mo/Mo combination and from 26 to 34kVp for the Mo/Rh combination. A simple model to describe the variation of TLD response R with energy E was derived. Thin-film TLDs on 2- to 7-cm breast thicknesses of BR-12 phantoms and an ACR accredited mammographic phantom were examined under automatic exposure control. Visibility of thin-film TLDs in digital mammographic images was assessed through subjective and objective methods. Entrance surface doses (ESDs) were measured and mean glandular doses (MGDs) were estimated. Results: The energy responses of GR-200F TLDs relative to Cs-137 are from 0.99 to 1.14, and increase slightly with the increase of mean energy. The TLD response R with energy E (keV) in this study is R(E)= {1-exp[-6.03 x 10(-3) x (E+ 1.17)1)]}[1+7.83 x 10(4) x (E +76.7)(-3)]. The digital images with the thin-film TLDs were only just visible on the phantom thickness less than 4cm. while the thin-film TLD was invisible on a 6cm thick phantom. The tube current-time products, ESDs and MGDs with and without thin-film TLDs under automatic exposure control were the same (P > 0.05). Conclusions: The energy responses increase slightly with the increase of mammographic energy. Thin-film TLDs are nearly invisible in digital mammograms. Furthermore. they neither affect the technical factor selection by aligning the equation on the text line automatic exposure control, nor change ESDs and MGDs. Hence, thin-film TLDs are suitable in vivo dosimetric measurements of mammography procedures.
机译:目的:评估乳腺摄影中薄膜TLD的能量响应和可见性。材料和方法:研究了钼-钼组合在24至32kVp,钼/铑组合在26-34kVp的乳腺摄影能量下GR-200F薄膜TLD的能量响应。推导了一个简单的模型来描述TLD响应R随能量E的变化。在自动曝光控制下,检查胸部厚度为2至7厘米的BR-12体模和ACR认可的乳腺摄影体模的薄膜TLD。通过主观和客观方法评估了数字X线摄影图像中薄膜TLD的可见性。测量入口表面剂量(ESD)并估算平均腺体剂量(MGD)。结果:相对于Cs-137,GR-200F TLD的能量响应为0.99至1.14,并且随着平均能量的增加而略有增加。在这项研究中,具有能量E(keV)的TLD响应R为R(E)= {1-exp [-6.03 x 10(-3)x(E + 1.17)1)]} [1 + 7.83 x 10(4) x(E +76.7)(-3)]。带有薄膜TLD的数字图像仅在幻影厚度小于4厘米时才可见。而薄膜TLD在6厘米厚的幻影上看不见。在自动曝光控制下,有和没有薄膜TLD的电子管电流乘积,ESD和MGD相同(P> 0.05)。结论:随着乳房X线照相能量的增加,能量反应略有增加。薄膜TLD在数字乳房X线照片中几乎是看不见的。此外。它们既不会通过在文本行自动曝光控制上对齐方程来影响技术因素的选择,也不会更改ESD和MGD。因此,薄膜TLD适用于乳腺X线摄影程序的体内剂量测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号