首页> 外文期刊>Radiation measurements >Performance of the Harshaw DXT-RAD (TLD-100) dosemeter
【24h】

Performance of the Harshaw DXT-RAD (TLD-100) dosemeter

机译:Harshaw DXT-RAD(TLD-100)剂量计的性能

获取原文
获取原文并翻译 | 示例
           

摘要

The Defence Science and Technology Laboratory (Dstl) has type-tested the Harshaw DXT-RAD dosemeter against the International Standard ISO 12794:2000(E) 'Individual thermoluminescence dosemeters for extremities and eyes' [International Organization for Standardization, 2000. International Standard ISO 12794:2000(E). Nuclear energy-radiation protection-individual thermoluminescence dosemeters for extremities and eyes]. The tests carried out included batch homogeneity, reproducibility, linearity, detection threshold, environmental stability (temperature, humidity and light), self irradiation, isotropy, energy response, dose-rate response and neutron response. The DXT-RAD dosemeter satisfied all the requirements of ISO 12794:2000(E), with the exception of its response at low photon energies and intermediate beta energies. However, it should be noted that other forms of the DXT-RAD dosemeter with a thinner entrance window are available to give an improved response to low-energy radiations.
机译:国防科学技术实验室(Dstl)已对Harshaw DXT-RAD剂量计进行了国际标准ISO 12794:2000(E)“肢体和眼睛的单独热致发光剂量计”的型式测试[国际标准化组织,2000年。国际标准ISO 12794:2000(E)。肢体和眼睛的核能辐射防护-个体热致发光剂量仪]进行的测试包括批处理的均质性,可重复性,线性,检测阈值,环境稳定性(温度,湿度和光),自辐射,各向同性,能量响应,剂量率响应和中子响应。 DXT-RAD剂量计满足ISO 12794:2000(E)的所有要求,但它在低光子能量和中等β能量下的响应除外。但是,应注意,其他形式的DXT-RAD剂量计也可以使用更薄的入射窗,以改善对低能辐射的响应。