首页> 外文期刊>Radiation measurements >An X-ray imaging detector based on pixel structured scintillator
【24h】

An X-ray imaging detector based on pixel structured scintillator

机译:基于像素结构闪烁体的X射线成像探测器

获取原文
获取原文并翻译 | 示例

摘要

An X-ray imaging detector consisting of a silicon charge coupled device (CCD) and pixel structured scintillator was developed to improve spatial resolution. The pixelated structure reduces spreading of scintillation light through the scintillator by providing air gap between scintillator pixels, and increased spatial resolution by reducing the cross-talk between CCD pixels. The modulation transfer Functions (MTF) of the non-pixelated and pixilated scintillator detector were compared. The improvement in spatial resolution was evaluated by both computational simulation and experimental measurement of MTF; both results showed coincidence. (c) 2007 Elsevier Ltd. All rights reserved.
机译:开发了由硅电荷耦合器件(CCD)和像素结构闪烁体组成的X射线成像检测器,以提高空间分辨率。像素化结构通过在闪烁体像素之间提供空气间隙来减少闪烁光通过闪烁体的扩散,并通过减少CCD像素之间的串扰来提高空间分辨率。比较了非像素化和像素化闪烁体检测器的调制传递函数(MTF)。通过计算仿真和MTF的实验测量来评估空间分辨率的提高;两种结果都符合。 (c)2007 Elsevier Ltd.保留所有权利。