Seven Ventilators Challenged With Leaks During Neonatal Nasal CPAP: An Experimental Pilot Study

机译:七名呼吸机在新生儿鼻CPAP期间面临泄漏的挑战:一项实验性试验研究

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Nasal CPAP is the most common respiratory support for neonates. Several factors are considered important for effective treatment, including leaks at the patient interface and the delivery of pressure-stable CPAP. Investigations of pressure stability during leaks should include both the change in the mean delivered CPAP and the pressure variation during each breath. The aim of this study was to examine the response of ventilators delivering nasal CPAP when challenged with leaks at the patient interface. METHODS: Seven ventilators providing nasal CPAP at 4 cmH(2)O were challenged with leaks during simulated neonatal breathing. Leak was applied for 15 consecutive breaths at a constant level (1-4 L/min). RESULTS: The 2 aspects of pressure stability were evaluated by measuring the mean delivered CPAP and the amplitude of pressure swings before, during, and after leaks. The ability to maintain the delivered CPAP and the amplitude of pressure swings varied greatly among the 7 ventilators before, during, and after leaks. Four of the ventilators tested have built-in leak compensation. CONCLUSIONS: There was no simple relationship between maintaining delivered CPAP during leaks and providing CPAP with low pressure swing amplitude. Maintaining the delivered CPAP and providing this without pressure swings are 2 separate aspects of pressure stability, and investigations concerning the clinical importance of pressure stability should address both aspects. This study also shows that compensation for leaks does not necessarily provide pressure-stable CPAP.
机译:背景:鼻CPAP是新生儿最常见的呼吸支持。几个因素被认为对有效治疗很重要,包括患者界面处的泄漏和压力稳定的CPAP的输送。对泄漏期间压力稳定性的研究应既包括平均输送CPAP的变化,也应包括每次呼吸期间的压力变化。这项研究的目的是检查在面对患者接口泄漏的情况下呼吸机输送鼻腔CPAP的反应。方法:在模拟新生儿呼吸过程中,有7台呼吸机在4 cmH(2)O处提供鼻CPAP,并面临泄漏挑战。以恒定水平(1-4 L / min)连续15次呼吸进行渗漏。结果:通过测量泄漏前,泄漏中和泄漏后的平均CPAP和压力波动幅度评估了压力稳定性的两个方面。在泄漏前,泄漏中和泄漏后的7台呼吸机中,维持输送的CPAP和压力波动幅度的能力差异很大。测试的四台呼吸机具有内置的泄漏补偿功能。结论:在泄漏期间维持输送的CPAP与为CPAP提供低压力摆幅之间没有简单的关系。保持所提供的CPAP并提供无压力波动的方法是压力稳定性的两个独立方面,有关压力稳定性临床重要性的研究应同时解决这两个方面。这项研究还表明,对泄漏的补偿并不一定能提供压力稳定的CPAP。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号