A comparison of the end-tidal CO2 measured by portable capnometer and the arterial PCO2 in spontaneously breathing patients.

机译:在自发呼吸患者中通过便携式二氧化碳监测仪测量的潮气末二氧化碳与动脉PCO2的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

An end-tidal CO2 (ETCO2) monitor (capnometer) is used most often as a noninvasive substitute for PaCO2 in anesthesia, anesthetic recovery and intensive care. However, the utility and accuracy of the portable capnometer in spontaneously breathing patients with or without chronic pulmonary diseases has received little recognition. To determine the utility of the portable capnometer in general wards and in in-home care, we examined the correlation between ETCO2 measured by a portable capnometer and simultaneous PaCO2 measured in 41 spontaneously breathing patients. TV-ETCO2 (ETCO2 measured by tidal volume maneuver) was lower than PaCO2 by an average of 9.0 mmHg and VC-ETCO2 (ETCO2 measured by vital capacity maneuver) was lower than PaCO2 by an average of 0.5 mmHg. The mean difference between PaCO2 and VC-ETCO2 was not statistically significant. Regression analysis showed a close correlation between VC-ETCO2 and PaCO2 (r = 0.91, P < 0.0001). Thus,VC-ETCO2 was highly correlated with PaCO2. Furthermore, a close correlation between VC-ETCO2 and PaCO2 was also observed in patients with compromised pulmonary function (r = 0.88, P < 0.0001 in patients with below 70% of FEV(1.0)%; r = 0.89, P < 0.0001 in patients with below 80% of %VC). Our studies show that VC-ETCO2 measured by the portable capnometer gives a reliable pointestimate of PaCO2, and can be useful to evaluate the respiratory condition of spontaneously breathing patients in general wards and in in-home care.
机译:在麻醉,麻醉恢复和重症监护中,潮气末二氧化碳(ETCO2)监测仪(二氧化碳测定仪)最常用作PaCO2的无创替代品。但是,便携式二氧化碳监测仪在有或没有慢性肺部疾病的自发呼吸患者中的实用性和准确性尚未得到认可。为了确定便携式二氧化碳监测仪在普通病房和家庭护理中的实用性,我们检查了由便携式二氧化碳监测仪测量的ETCO2与41例自发呼吸患者同时测量的PaCO2之间的相关性。 TV-ETCO2(通过潮气量测得的ETCO2)比PaCO2低9.0 mmHg,而VC-ETCO2(通过肺活量测得的ETCO2)比PaCO2低0.5 mmHg。 PaCO2和VC-ETCO2之间的平均差异无统计学意义。回归分析显示VC-ETCO2和PaCO2之间密切相关(r = 0.91,P <0.0001)。因此,VC-ETCO2与PaCO2高度相关。此外,在肺功能受损的患者中也观察到VC-ETCO2与PaCO2之间的密切相关(FEV(1.0)%低于70%的患者r = 0.88,P <0.0001; FEV低于70%的患者r = 0.89,P <0.0001低于%VC的80%)。我们的研究表明,通过便携式二氧化碳监测仪测量的VC-ETCO2给出了PaCO2的可靠估计值,可用于评估普通病房和家庭护理中自发呼吸患者的呼吸状况。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号