摘要

Purpose: To describe the mechanical failure of a 25-gauge vitreous cutter intra-operatively and the management of this complication.Methods: An observational case report of a 39-year-old Type 1 diabetic woman with proliferative retinopathy who underwent pars plana vitrectomy with 25-gauge instrumentation.Results: Breakage of the 25-gauge vitreous cutter occurred during delamination of a dense fibrovascular plaque. The cutter was successfully removed via the trocar as a single unit. The patient recovered with no sequelae.Conclusions: Breakage of 25-gauge instrumentation intraoperatively is a very rare event. This case constitutes only the second reported complication as such. However, it is easily managed and can result in few, if any, adverse sequelae.
机译:目的:描述25号玻璃体切割器的术中机械故障及并发症的处理方法:观察性报道一名39岁的1型糖尿病性增生性视网膜病变妇女接受了全玻璃体玻璃体切除术25号仪器。结果:25号玻璃体切割器破裂发生在致密的纤维血管斑块分层时。切刀已通过套管针作为一个整体成功移除。患者康复后无后遗症。结论:术中25号仪表的损坏是非常罕见的事件。这种情况仅构成第二例报告的并发症。但是,它很容易管理,几乎不会产生不良后遗症。