Quality control of the stereotactic radiosurgery procedure with the polymer-gel dosimetry.

机译:聚合物凝胶剂量测定法对立体定向放射外科手术程序的质量控制。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To assess the entire geometric and dosimetric (relative) uncertainties of the radiosurgery procedure with the Leksell gamma knife. MATERIALS AND METHODS: The entire Leksell gamma knife stereotactic radiosurgery treatment procedure was simulated with the use of a special water filled head phantom and polymer-gel dosimeter evaluated by nuclear magnetic resonance (NMR). A test vessel filled with the polymer-gel dosimeter was fixed in the head phantom. The phantom underwent stereotactic NMR imaging, treatment planning and then irradiation according to the treatment plan prepared exactly the same way as in the ordinary treatment procedure for a patient. The treatment plan was represented by one isocenter positioned approximately centrally in the head phantom. This procedure was subsequently repeated for all four collimators (4, 8, 14, 18mm) used on the Leksell gamma knife. Evaluation of dosimeters was performed on a Siemens EXPERT 1T NMR scanner. Dose profiles in X, Y and Z axes through the ellipsoidal shaped dose distribution were obtained to compare experimental results from the irradiated phantom with the treatment planning system calculations. RESULTS: Reasonable agreement was observed between the treatment planning system calculations of relative dose distribution and the measured data. The maximum observed deviation in the spatial position between the center of the measured and calculated dose profiles was 0.6mm. The maximum observed difference in full width of half maximum between calculated and measured profiles was 1.2mm. CONCLUSIONS: The use of polymer-gel dosimetry for a verification of stereotactic procedures has some unique advantages that can be summarized as follows: the dosimeter itself is tissue equivalent, three-dimensional dose distributions can be measured and the dosimeter allows simulation of the patient's procedures without any limitations.
机译:目的:评估用Leksell伽马刀进行放射外科手术的整个几何和剂量学(相对)不确定性。材料和方法:整个Leksell伽玛刀立体定向放射外科手术治疗过程是通过使用特殊的充满水的头部模型和通过核磁共振(NMR)评估的聚合物凝胶剂量计来模拟的。将装有聚合物-凝胶剂量计的测试容器固定在头部模型中。对体模进行了立体定向NMR成像,治疗计划,然后根据与常规治疗程序完全相同的方式为患者准备的治疗计划进行照射。该治疗计划以一个大约位于头模中心的等角点表示。随后,对Leksell伽玛刀上使用的所有四个准直仪(4、8、14、18mm)重复此过程。剂量计的评估是在Siemens EXPERT 1T NMR扫描仪上进行的。获得了通过椭圆形剂量分布在X,Y和Z轴上的剂量分布,以比较受照体模的实验结果与治疗计划系统的计算结果。结果:在治疗计划系统相对剂量分布的计算与测量数据之间观察到合理的一致性。测量和计算的剂量分布图中心之间的空间位置的最大观察到的偏差为0.6mm。在计算出的轮廓和测量的轮廓之间的最大半宽度的最大观察到的差异为1.2mm。结论:使用聚合物凝胶剂量测定法来验证立体定向手术方法具有一些独特的优势,可以总结如下:剂量计本身是组织等效的,可以测量三维剂量分布,并且剂量计可以模拟患者的手术程序没有任何限制。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号