This HIPAA-compliant study had institutional review board approval, with waiver of informed consent. The purpose was to test the hypothesis that, compared with a standard magnetic resonance (MR) cholangiography sequence, MR cholangiography with a sampling perfection with application optimized contrasts using different flip angle evolutions (SPACE) sequence reduces ghosting artifacts while maintaining image quality and sufficient contrast-to-noise ratio (CNR) at 3.0 T. The study population consisted of 15 women and 14 men (mean age, 47.2 years +/- 18.1 [standard deviation]) who were consecutively referred for MR cholangiography between November 2004 and November 2005. Acquisition times were lower and N/2 ghosting artifacts were eliminated with SPACE. However, the SPACE sequence yielded images that were visually grainier and had lower CNR. Overall, the readers preferred the appearance of images obtained with the SPACE sequence, most likely because of the elimination of N/2 ghosting artifacts.
机译:此符合HIPAA的研究已获得机构审查委员会的批准,并放弃了知情同意。目的是检验以下假设:与标准磁共振(MR)胆管造影术序列相比,具有不同采样率的MR胆管造影术具有使用不同翻转角演变(SPACE)序列进行了应用优化的对比,可减少重影伪影,同时保持图像质量和足够的对比度噪声比(CNR)在3.0T。研究人群包括15位女性和14位男性(平均年龄47.2岁+/- 18.1 [标准差]),这些患者在2004年11月至2005年11月期间连续接受MR胆管造影采集时间更短,并且使用SPACE消除了N / 2个重影伪影。但是,SPACE序列产生的图像在视觉上较粗糙,且CNR较低。总体而言,读者更喜欢使用SPACE序列获得图像的外观,这很可能是因为消除了N / 2重影伪影。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号