Artifacts caused by insufficient contrast medium filling during C-arm cone-beam CT scans: a phantom study.

机译:在C型臂锥束CT扫描过程中,由于造影剂填充不足而导致的伪影:一项幻像研究。

获取原文
获取原文并翻译 | 示例
           

摘要

We investigated artifacts due to late-arriving contrast medium (CM) during C-arm cone-beam computed tomography. We scanned a phantom filled with water or with 100, 50, or 5% v/v concentrations of CM and then virtually produced CM-delayed projection data by partially replacing the projection images. Artifacts as a function of concentration, percentage of filling time, and size and position of the filling area were assessed. In addition, we used an automatic power injector with different injection delays to inject CM during the scans. A decrease in filling times caused by a lag in CM arrival during the scan resulted in a decrease in pixel values, distortion of the filling area, and appearance of streak artifacts. Even a delay of approximately 20% in CM arrival in the total scan time resulted in obvious distortion of the filling area. The distortion and streak artifacts tended to worsen at higher CM concentrations. Use of a minimum CM concentration based on the purpose of the examination and constant filling at the target region are effective for avoiding these artifacts.
机译:我们调查了由于C型臂锥束计算机断层扫描中迟到的造影剂(CM)造成的伪影。我们扫描了装满水或100、50或5%v / v浓度的CM的幻像,然后通过部分替换投影图像来虚拟生成CM延迟的投影数据。评估伪影随浓度,填充时间百分比以及填充区域的大小和位置的变化。此外,我们在扫描过程中使用了具有不同注入延迟的自动动力注射器来注入CM。在扫描期间由于CM到达延迟而导致的填充时间减少,导致像素值减少,填充区域变形以及出现条纹伪影。甚至在整个扫描时间内CM到达大约20%的延迟也会导致填充区域明显变形。在较高的CM浓度下,失真和条纹伪影趋于恶化。根据检查的目的使用最小的CM浓度并在目标区域恒定填充对于避免这些伪影是有效的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号