Fat-subtracted three-dimensional time-of-flight MR angiography of the neck by use of fat-only images with the two-point Dixon technique

机译:通过使用仅脂肪图像和两点Dixon技术,对颈部进行脂肪减去三维飞行时间MR血管造影

获取原文
获取原文并翻译 | 示例
           

摘要

For improvement of three-dimensional time-of-flight magnetic resonance angiography (3D-TOF-MRA) image quality in the neck, fat-subtracted MRA by use of the two-point Dixon technique was compared with conventional fat-suppressed MRA techniques. Three different types of neck 3D-TOF-MRA were obtained [minimum echo time (TE) (1.9 ms), opposed-phase TE (3.4 ms), and chemical shift selective fat suppression (CHESS) (TE = 1.9 ms)] on five volunteers at 3.0 T. MRA was obtained with subtraction of fat-only images (produced by a two-point Dixon sequence) from minimum-TE MRA images, and compared with other fat-suppressed MRA images. Fat-subtracted MRA demonstrated uniform fat suppression compared with other techniques. The mean vessel-to-fat contrast in fat-subtracted MRA was significantly higher (p < 0.01) than in other MRA images (minimum-TE: 0.137 ± 0.086, opposed-phase TE: 0.268 ± 0.102, CHESS: 0.307 ± 0.052, fat-subtracted: 0.965 ± 0.101). The mean vessel-to-muscle contrast in opposed-phase TE MRA was significantly lower (p < 0.01) than in other MRA images (minimum-TE: 0.526 ± 0.036, opposed-phase TE: 0.419 ± 0.188, CHESS: 0.511 ± 0.023, fat-subtracted: 0.573 ± 0.016). Fat-subtracted MRA by use of the two-point Dixon technique improves the image quality of neck MRA. This technique would be a useful method for MRA, especially in areas with inhomogeneous magnetic fields, such as the neck.
机译:为了改善脖子上的三维飞行时间磁共振血管造影(3D-TOF-MRA)图像质量,将使用两点Dixon技术的减脂MRA与常规减脂MRA技术进行了比较。在三种情况下获得了三种不同类型的颈部3D-TOF-MRA [最小回声时间(TE)(1.9 ms),反相TE(3.4 ms)和化学位移选择性脂肪抑制(CHESS)(TE = 1.9 ms)]。五名3.0 T志愿者。MRA是通过从最小TE MRA图像中减去仅脂肪图像(由两点Dixon序列生成)而获得的,并与其他受脂肪抑制的MRA图像进行了比较。与其他技术相比,减脂MRA具有均匀的脂肪抑制作用。减去脂肪的MRA中的平均血管与脂肪对比显着高于(p <0.01)(最小TE:0.137±0.086,反相TE:0.268±0.102,CHESS:0.307±0.052,减去脂肪:0.965±0.101)。相对相TE MRA的平均血管与肌肉对比显着低于(p <0.01)(最低TE:0.526±0.036,相对TE:0.419±0.188,CHESS:0.511±0.023 ,减去脂肪:0.573±0.016)。通过使用两点Dixon技术减去脂肪的MRA可以改善颈部MRA的图像质量。对于MRA,该技术将是一种有用的方法,尤其是在具有不均匀磁场的区域(例如颈部)中。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号