首页> 外文期刊>Radiological physics and technology >Power Doppler signal calibration between ultrasound machines by use of a capillary-flow phantom for pannus vascularity in rheumatoid finger joints: a basic study
【24h】

Power Doppler signal calibration between ultrasound machines by use of a capillary-flow phantom for pannus vascularity in rheumatoid finger joints: a basic study

机译:基于毛细管流幻影的类风湿手指关节pan血管的超声多普勒机之间功率多普勒信号校准:基础研究

获取原文
获取原文并翻译 | 示例
           

摘要

Ultrasound allows the detection and grading of inflammation in rheumatology. Despite these advantages of ultrasound in the management of rheumatoid patients, it is well known that there are significant machine-to-machine disagreements regarding signal quantification. In this study, we tried to calibrate the power Doppler (PD) signal of two models of ultrasound machines by using a capillary-flow phantom. After flow velocity analysis in the perfusion cartridge at various injection rates (0.1-0.5 ml/s), we measured the signal count in the perfusion cartridge at various injection rates and pulse repetition frequencies (PRFs) by using PD, perfusing an ultrasound micro-bubble contrast agent diluted with normal saline simulating human blood.
机译:超声可以检测和分级风湿病中的炎症。尽管超声在类风湿患者的治疗中具有这些优势,但是众所周知,在信号量化方面存在重大的机器对机器分歧。在这项研究中,我们尝试通过使用毛细管流动体模来校准两种型号的超声仪的功率多普勒(PD)信号。在以不同的注入速率(0.1-0.5 ml / s)在灌注盒中进行流速分析后,我们通过使用PD,灌注超声微测量仪在不同的注入速率和脉冲重复频率(PRF)下测量了灌注盒中的信号计数。用生理盐水稀释的气泡造影剂,模拟人的血液。