首页> 外文期刊>Radiological physics and technology >Validation of a quick three-dimensional dose verification system for pre-treatment IMRT QA
【24h】

Validation of a quick three-dimensional dose verification system for pre-treatment IMRT QA

机译:用于治疗前IMRT QA的快速三维剂量验证系统的验证

获取原文
获取原文并翻译 | 示例
           

摘要

In this study, we evaluated the dosimetric performance of the three-dimensional (3D) dose verification system, COMPASS version 3 (IBA Dosimetry, GmbH, Germany). The COMPASS has the function of a dedicated beam modeling and dose calculation. It is able to reconstruct 3D dose distributions on patient CT images, using the incident fluence from a linear accelerator measured with the MatriXX 2D array (IBA Dosimetry). The dose profiles measured with various multi-leaf collimator (MLC) test patterns for the COMPASS were checked by comparison with those of EDR2 (Eastman Kodak, Rochester, NY) films and Monte Carlo (MC) simulations. The COMPASS was also used for dose verification in clinical intensity-modulated radiation therapy (IMRT) plans for head and neck cases.
机译:在这项研究中,我们评估了三维(3D)剂量验证系统COMPASS版本3(IBA Dosimetry,GmbH,Germany)的剂量学性能。 COMPASS具有专用的射束建模和剂量计算功能。它能够通过使用MatriXX 2D阵列(IBA剂量测定法)测量的线性加速器的入射通量来重建患者CT图像上的3D剂量分布。通过与EDR2(Eastman Kodak,Rochester,NY)胶片和Monte Carlo(MC)模拟的比较,检查了用COMPASS的各种多叶准直器(MLC)测试模式测量的剂量分布。 COMPASS还用于头颈部病例的临床强度调制放射治疗(IMRT)计划中的剂量验证。