首页> 外文期刊>Radiologic Technology >Radiation dose in pelvic imaging.
【24h】

Radiation dose in pelvic imaging.

机译:骨盆成像中的辐射剂量。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To compare the radiation dose during pelvic x-ray examinations using computed radiography (CR) and film-screen (FS) radiography at various x-ray tube voltages (kV) and tube-current time product (mAs) values. METHODS: A pelvic phantom was imaged using FS and CR systems. The entrance surface dose was measured using an ionization chamber, and the gonadal dose and effective dose were calculated using the XDOSE program. The diagnostic quality of the images was assessed using a 5-point subjective scoring system. RESULTS: At standard kV values, the image quality did not vary significantly between the CR and the FS system, but at higher kV values, the CR images werefound to be of better quality than FS images. In addition, the lower limit of entrance skin dose consistent with diagnostically acceptable CR images was 50% lower than that for FS images. CONCLUSION: The gonadal dose and effective dose for pelvic x-ray examinations can be reduced by 50% when CR systems are used and appropriate exposure factors are established.
机译:目的:比较在各种X射线管电压(kV)和管电流时间乘积(mAs)值下使用计算机射线照相(CR)和胶片屏幕(FS)射线照相进行的骨盆X线检查期间的辐射剂量。方法:使用FS和CR系统对骨盆体模进行成像。使用电离室测量入射表面剂量,并使用XDOSE程序计算性腺剂量和有效剂量。使用5分主观评分系统评估图像的诊断质量。结果:在标准kV值下,CR和FS系统之间的图像质量没有明显变化,但是在更高kV值下,发现CR图像的质量比FS图像更好。另外,与诊断上可接受的CR图像一致的进入皮肤剂量的下限比FS图像低50%。结论:使用CR系统并建立适当的暴露因素后,用于盆腔X线检查的性腺剂量和有效剂量可减少50%。