摘要

The quality control of liquid-crystal display (LCD) monitors has become one of the important topics for maintaining reliable soft-copy readings in the interpretation of diagnostic images. In this paper, the effects of correction in the luminance measurement of an LCD monitor by use of a telescopic-type luminance meter were investigated. The luminance of the LCD monitor in different ambient-lighting conditions was measured and compared to the results obtained with no ambient lighting (0 lux). The reproducibility of luminance measurements and luminance ratios without a baffled tube was lower than those measured with the baffled tube due to the effect of ambient light. These tendencies were obvious at a relatively low luminance. The correction method by subtraction of the reflected ambient light on the surface of the LCD monitor and the stray light of the telescopic-type luminance meter from the measured luminance was examined. We found that the correction was able to bring the luminance close to that measured with the baffled tube.
机译:液晶显示器(LCD)监视器的质量控制已成为在诊断图像的解释中保持可靠的软拷贝读数的重要主题之一。在本文中,研究了使用伸缩式亮度计在LCD监视器的亮度测量中进行校正的效果。测量了在不同环境照明条件下LCD监视器的亮度,并将其与没有环境照明(0 lux)的结果进行了比较。由于环境光的影响,没有挡板管的亮度测量结果和亮度比的可重复性低于使用挡板管测量的结果。这些趋势在较低的亮度下很明显。研究了通过从测量的亮度中减去LCD监视器的表面上的反射环境光和伸缩型亮度计的杂散光的校正方法。我们发现该校正能够使亮度接近使用挡板管测得的亮度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号