Monte Carlo simulations of electron beams collimated with a dual electron multileaf collimator: a feasibility study.

机译:用双电子多叶准直仪准直的电子束的蒙特卡罗模拟:可行性研究。

获取原文
获取原文并翻译 | 示例
           

摘要

Electron applicators and cut-outs have been used for some time in the delivery of electron beam therapy. A dual electron multileaf collimator (eMLC) consisting of upper and lower eMLCs was designed, and dose distributions of electron beams delivered by it were evaluated by Monte Carlo simulations by use of the EGSnrc Monte Carlo code. The percentage depth dose (PDD), dose profiles, dose gradient falloff (G), depth of maximum dose (R(100)), surface dose, bremsstrahlung background, beam flatness, and penumbra of the dual eMLC were evaluated and compared with those simulated and measured with the standard applicators inserted into the treatment head of the medical linear accelerator (linac). The results showed good agreement in most cases. Specifically, the flatness and penumbra obtained with the dual eMLC were better than those obtained with standard applicators. It is therefore possible to use the proposed dual eMLC in the delivery of electron beam therapy without the need for applicators and cut-outs. This will minimize the inconvenience of placing cut-outs on the patients; also, changes in the required field can be effected without the therapist going into the accelerator room when the dual eMLC is finally automated. The absence of a helium bag between the upper and lower eMLCs did not offer significant changes in the dosimetry parameters compared to the eMLC fitted with the helium bag.
机译:在电子束治疗的交付中,电子施药器和切口已经使用了一段时间。设计了由上,下eMLC组成的双电子多叶准直仪(eMLC),并使用EGSnrc蒙特卡洛代码通过蒙特卡洛模拟对由其发射的电子束的剂量分布进行了评估。对双重eMLC的百分比深度剂量(PDD),剂量分布,剂量梯度下降(G),最大剂量深度(R(100)),表面剂量,致背景,光束平坦度和半影进行了评估并与之进行了比较使用插入到医用线性加速器(直线加速器)治疗头中的标准涂药器进行仿真和测量。在大多数情况下,结果显示出良好的一致性。特别地,用双重eMLC获得的平坦度和半影区比使用标准涂药器获得的平坦度和半影区更好。因此,有可能在电子束治疗的递送中使用所提出的双重eMLC,而无需施加器和切口。这将最小化在患者身上放置切口的不便;同样,当双重eMLC最终实现自动化时,无需治疗师进入加速器室即可进行必填字段的更改。与装有氦气袋的eMLC相比,上下eMLC之间没有氦气袋不会在剂量学参数上产生重大变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号