首页> 外文期刊>Radiologic Technology >Variation in CR imaging plate readers.
【24h】

Variation in CR imaging plate readers.

机译:CR成像板读取器的变化。

获取原文
获取原文并翻译 | 示例
           

摘要

This study investigated computed radiography (CR) image processing to determine whether variation exists within and among CR imaging plate readers. Photostimulable imaging plates were exposed using a phantom test tool and processed in 4 CR readers located in diverse settings in an urban academic medical center. Research results indicate daily variation of S-numbers within individual CR readers did not exceed tolerance limits, although over the 3-week study period, evidence of S-number variation within individual CR readers was mixed. In addition, S-number variability among multiple CR readers was found to be statistically significant. Although the cause of the variability remains unknown, evidence of variability among multiple CR readers indicates a need for systematic quality control.
机译:这项研究调查了计算机射线照相(CR)图像处理,以确定CR成像板读取器内部和之间是否存在差异。使用体模测试工具对光敏成像板进行曝光,并在位于城市学术医疗中心不同环境的4个CR读取器中对其进行处理。研究结果表明,尽管在3周的研究期内,单个CR阅读器内S值的每日变化没有超出公差极限,但证据却不尽相同。此外,发现多个CR阅读器之间的S数差异具有统计学意义。尽管变化的原因仍然未知,但多个CR读取器之间变化的证据表明需要系统的质量控制。