首页> 外文期刊>Luminescence: The journal of biological and chemical luminescence >A novel green analytical procedure for monitoring of azoxystrobin in water samples by a flow injection chemiluminescence method with off-line ultrasonic treatment
【24h】

A novel green analytical procedure for monitoring of azoxystrobin in water samples by a flow injection chemiluminescence method with off-line ultrasonic treatment

机译:离线超声处理流动注射化学发光法监测水样中嘧菌酯的新型绿色分析程序

获取原文
获取原文并翻译 | 示例

摘要

A simple and green flow injection chemiluminescence (FI-CL) method for determination of the fungicide azoxystrobin was described for the first time. CL signal was generated when azoxystrobin was injected into a mixed stream of luminol and KMnO_4. The CL signal of azoxystrobin could be greatly improved when an off-line ultrasonic treatment was adopted. Meanwhile, the signal intensity increases with the analyte concentration proportionally. Several variables, such as the ultrasonic parameters, flow rate of reagents, concentrations of sodium hydroxide solution and CL reagents (potassium permanganate, luminol) were investigated, and the optimal CL conditions were obtained. Under optimal conditions, the linear range of 1-100 ng/mL for azoxystrobin was obtained and the detection limit (3σ) was determined as 0.13 ng/mL. The relative standard deviation was 1.5% for 10 consecutive measurements of 20 ng/mL azoxystrobin. The method has been applied to the determination of azoxystrobin residues in water samples.
机译:首次介绍了一种简单而绿色的流动注射化学发光(FI-CL)方法来测定杀真菌剂嘧菌酯。当将天青霉素添加到鲁米诺和KMnO_4的混合流中时,会产生CL信号。采用离线超声处理可以显着提高嘧菌酯的CL信号。同时,信号强度随分析物浓度成比例增加。研究了超声波参数,试剂流速,氢氧化钠溶液浓度和CL试剂(高锰酸钾,鲁米诺)等变量,确定了最佳的CL条件。在最佳条件下,对嘧菌酯的线性范围为1-100 ng / mL,检测限(3σ)确定为0.13 ng / mL。连续10次测量20 ng / mL嘧菌酯的相对标准偏差为1.5%。该方法已用于测定水样中的嘧菌酯残留。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号