首页> 外文期刊>NMR in biomedicine >Uniform prostate imaging and spectroscopy at 7 T: Comparison between a microstrip array and an endorectal coil
【24h】

Uniform prostate imaging and spectroscopy at 7 T: Comparison between a microstrip array and an endorectal coil

机译:7 T时的均匀前列腺成像和光谱:微带阵列和直肠内线圈之间的比较

获取原文
获取原文并翻译 | 示例
           

摘要

An endorectal coil and an eight-element microstrip array were compared for prostate imaging at 7 T. An extensive radiofrequency safety assessment was performed with the use of finite difference time domain simulations to determine safe scan parameters. These simulations showed that the endorectal coil can deliver substantially more B1+ to the prostate than can the microstrip array within the specific absorption rate safety guidelines. However, the B1+ field of the endorectal coil is very inhomogeneous, which makes the use of adiabatic pulses compulsory for T_1- or T_2-weighted imaging. As a consequence, a full prostate examination is only possible in a feasible amount of time when the microstrip array is used for T_1- and T_2-weighted imaging, whereas the endorectal coil is required for spectroscopic imaging. The pulse parameters were optimised within the specific absorption rate guidelines and thereafter used to provide a good illustration of the possibilities of prostate imaging at 7 T.
机译:比较了直肠内线圈和八元素微带阵列在7 T时的前列腺成像。使用有限差分时域模拟进行了广泛的射频安全性评估,以确定安全扫描参数。这些模拟表明,在特定吸收率安全性准则内,直肠内线圈比微带阵列可以向前列腺输送的B1 +明显更多。但是,直肠内线圈的B1 +场非常不均匀,这使得必须对T_1或T_2加权成像使用绝热脉冲。结果,当微带阵列用于T_1和T_2加权成像时,仅在可行的时间内进行完整的前列腺检查,而光谱成像则需要直肠内线圈。脉冲参数在特定的吸收率指南内进行了优化,然后用于很好地说明7 T时前列腺成像的可能性。