首页> 外文期刊>Lab on a chip >Blood coagulation screening using a paper-based microfluidic lateral flow device
【24h】

Blood coagulation screening using a paper-based microfluidic lateral flow device

机译:使用纸基微流控侧向流动装置进行凝血检查

获取原文
获取原文并翻译 | 示例
           

摘要

A simple approach to the evaluation of blood coagulation using a microfluidic paper-based lateral flow assay (LFA) device for point-of-care (POC) and self-monitoring screening is reported. The device utilizes whole blood, without the need for prior separation of plasma from red blood cells (RBC). Experiments were performed using animal (rabbit) blood treated with trisodium citrate to prevent coagulation. CaCl2 solutions of varying concentrations are added to citrated blood, producing Ca~(2+) ions to re-establish the coagulation cascade and mimic different blood coagulation abilities in vitro. Blood samples are dispensed into a paper-based LFA device consisting of sample pad, analytical membrane and wicking pad. The porous nature of the cellulose membrane separates the aqueous plasma component from the large blood cells. Since the viscosity of blood changes with its coagulation ability, the distance RBCs travel in the membrane in a given time can be related to the blood clotting time. The distance of the RBC front is found to decrease linearly with increasing CaCl2 concentration, with a travel rate decreasing from 3.25 mm min~(-1) for no added CaCl2 to 2.2 mm min~(-1) for 500 mM solution. Compared to conventional plasma clotting analyzers, the LFA device is much simpler and it provides a significantly larger linear range of measurement. Using the red colour of RBCs as a visible marker, this approach can be utilized to produce a simple and clear indicator of whether the blood condition is within the appropriate range for the patient's condition.
机译:据报道,使用基于微流纸的侧向流动测定(LFA)装置进行即时护理(POC)和自我监测筛选的一种简单的评估血液凝固的方法。该设备利用全血,无需事先从红细胞(RBC)中分离血浆。使用经过柠檬酸三钠处理的动物(兔)血液进行实验以防止凝血。将不同浓度的CaCl2溶液添加到柠檬酸血中,产生Ca〜(2+)离子,以重新建立凝血级联并模仿体外的不同凝血能力。血液样本被分配到基于纸的LFA设备中,该设备由样本垫,分析膜和芯吸垫组成。纤维素膜的多孔性质将血浆血浆成分与大血细胞分开。由于血液的粘度随其凝结能力而变化,因此RBC在给定时间内在膜中的移动距离可能与血液凝固时间有关。发现RBC前缘的距离随CaCl2浓度的增加而线性减小,行进速率从不添加CaCl2的3.25 mm min〜(-1)降低到500 mM溶液的2.2 mm min〜(-1)。与传统的血浆凝结分析仪相比,LFA设备要简单得多,并且线性测量范围要大得多。使用RBC的红色作为可见标记,可以使用这种方法来生成简单清晰的指示符,以表明血液状况是否在适合患者状况的范围内。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号