An innovative sample-to-answer polymer lab-on-a-chip with on-chip reservoirs for the POCT of thyroid stimulating hormone (TSH)

机译:创新的从样品到答案的聚合物片上实验室,带有用于甲状腺刺激激素(TSH)POCT的片上储存器

获取原文
获取原文并翻译 | 示例
           

摘要

A new sample-to-answer polymer lab-on-a-chip, which can perform immunoassay with minimum user intervention through on-chip reservoirs for reagents and single-channel assay system, has been designed, developed and successfully characterized as a point-of-care testing (POCT) cartridge for the detection of thyroid stimulating hormone (TSH). Test results were obtained within 30 minutes after a sample was dropped into the POCT cartridge. The analyzed results of TSH showed a linear range of up to 55 μIU mL~(-1) with the limit of detection (LOD) of 1.9 μIU mL~(-1) at the signal-to-noise ratio (SNR) of 3. The reagents stored in the on-chip reservoirs maintained more than 97% of their initial volume for 120 days of storage time while the detection antibody retained its activity above 98% for 120 days. The sample-to-answer polymer lab-on-a-chip developed in this work using the mass-producible and low-cost polymer is well suited for the point-of-care testing of rapid in vitro diagnostics (IVD) of TSH.
机译:设计,开发并成功地设计了一种新型的从样品到答案的聚合物芯片实验室,该试剂可以通过最少的用户干预通过芯片上的试剂盒和单通道测定系统进行免疫测定,并成功地将其表征为:即时检测(POCT)药筒,用于检测甲状腺刺激激素(TSH)。在将样品放入POCT盒后的30分钟内获得测试结果。 TSH的分析结果显示线性范围高达55μIUmL〜(-1),检测限(LOD)为1.9μIUmL〜(-1),信噪比(SNR)为3存储在片上存储器中的试剂在120天的存储时间内保持了其初始体积的97%以上,而检测抗体在120天中仍保持其98%以上的活性。在这项工作中使用大规模生产且低成本的聚合物开发的“样品到答案的聚合物芯片实验室”非常适合于TSH的快速体外诊断(IVD)的即时检测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号