In order to assess the agreement of body weight (BW), body fat index values between total body (foot to hand) and segment body (foot to foot) analyzers, we measured BW and body fat percentage (BF%) of 37 subjects by both analyzers in the same condition. As a result, the Pearson correlation of body weight and BF% values were r(BW) = 1.00 and r(BF%) = 0.81, respectively (p < 0.0001). The mean difference of BW and BF% (total minus segment) were d(BW) = -0.281kg and d(BF%) = 0.662% respectively; and the standard deviation of these differences were S-d(BW) = 0.143 and S-d(BF%) = 3.898. In addition, the 95% limits of agreement of difference of BW was (-0.567 to +0.005) and (-7.135 to +8.459) for difference of BF%. In conclusion, the results indicated that agreement of BW and BF% values between the two analyzers is acceptable, thus they might be interchangeable for epidemiological studies.
机译:为了评估体重(BW),全身(脚到手)和分段身体(脚到脚)分析仪之间的体脂指数值的一致性,我们测量了37名受试者的BW和体脂百分比(BF%)两个分析仪处于相同状态。结果,体重和BF%值的皮尔逊相关性分别为r(BW)= 1.00和r(BF%)= 0.81(p <0.0001)。 BW和BF%(总负部分)的平均差分别为d(BW)= -0.281kg和d(BF%)= 0.662%;这些差异的标准偏差为S-d(BW)= 0.143和S-d(BF%)= 3.898。另外,BW差异的95%一致性限制为(-0.567至+0.005)和BF%差异的(-7.135至+8.459)。总之,结果表明两个分析仪之间的体重和BF%值的一致性是可以接受的,因此在流行病学研究中它们可能是可互换的。